Implantation of pipeline flow-diverting stents reduces aneurysm inflow without relevantly affecting static intra-aneurysmal pressure

Affiliations

Implantation of pipeline flow-diverting stents reduces aneurysm inflow without relevantly affecting static intra-aneurysmal pressure

Hans U Kerl et al. Neurosurgery. 2014 Mar.

Abstract

Background: Flow-diverting stent (FDS) implantation is an endovascular treatment option for intracranial aneurysms. However, little is known about the hemodynamic effects.

Objective: To assess the effect of stent compression on FDS porosity, to evaluate the influence of single and overlapping implantation of FDS on intra-aneurysmal flow profiles, and to correlate stent porosity with changes in static mean intra-aneurysmal pressure.

Methods: Intra-aneurysmal time-density curves were recorded in a pulsatile in vitro flow model before and after implantation of FDSs (Pipeline Embolization Device; ev3) in 7 different types of aneurysm models. Reductions in the maximum contrast inflow and time to maximum intra-aneurysmal contrast were calculated. Micro--computed tomography was performed, and compression-related FDS porosity was measured. The influence of FDS placement on mean static intra-aneurysmal pressure was measured.

Results: FDS compression resulted in an almost linear reduction in stent porosity. Stent porosity (struts per 1 mm) correlated significantly with the reduction of aneurysm contrast inflow (R = 0.81, P < .001) and delay until maximum contrast (R = 0.34, P = .001). Circulating intra-aneurysmal high-velocity flow was terminated in all sidewall models after implantation of a single stent. Superimposition of 2 stents reduced maximum intra-aneurysmal contrast by 69.1 ± 3.1% (mean ± SD) in narrow-necked sidewall aneurysm models, whereas no substantial reduction in maximum intra-aneurysmal contrast was observed in wide-necked sidewall aneurysm models. Intra-aneurysmal mean static pressure did not correlate with FDS porosity or number of implanted stents.

Conclusion: Implantation of FDS effectively reduces aneurysm inflow in a porosity-dependent way without relevantly affecting static mean intra-aneurysmal pressure.

Abbreviations: FDS, flow-diverting stentMAP, mean arterial pressurePED, Pipeline Embolization Device.