Low validity of the Sensewear Pro3 activity monitor compared to indirect calorimetry during simulated free living in patients with osteoarthritis of the hip
- PMID: 24552503
- PMCID: PMC3938645
- DOI: 10.1186/1471-2474-15-43
Low validity of the Sensewear Pro3 activity monitor compared to indirect calorimetry during simulated free living in patients with osteoarthritis of the hip
Abstract
Background: To validate physical activity estimates by the Sensewear Pro3 activity monitor compared with indirect calorimetry during simulated free living in patients diagnosed with osteoarthritis of the hip pre or post total hip arthroplasty.
Methods: Twenty patients diagnosed with hip osteoarthritis (10 pre- and 10 post total hip arthroplasty; 40% female; age: 63.3 ± 9.0; BMI: 23.7 ± 3.7). All patients completed a 2 hour protocol of simulated free living with 8 different typical physical activity types. Energy consumption (kcal/min) was estimated by the Sense Wear pro3 Armband activity monitor and validated against indirect calorimetry (criterion method) by means of a portable unit (Cosmed K4b2). Bias and variance was analyzed using functional ANOVA.
Results: Mean bias during all activities was 1.5 Kcal/min 95%CI [1.3; 1.8] corresponding to 72% (overestimation). Normal gait speed showed an overestimation of 2.8 Kcal/min, 95%CI [2.3; 3.3] (93%) while an underestimation of -1.1 Kcal/min, 95%CI [-1.8; -0.3] (-25%) was recorded during stair climb. Activities dominated by upper body movements showed large overestimation with 4.37 Kcal/min, 95%CI [3.8; 5.1] (170%) being recorded during gardening. Both bias and variance appeared to be dependent on activity type.
Conclusion: The activity monitor generally overestimated the energy consumption during common activities of low to medium intensity in the patient group. The size and direction of the bias was highly dependent on the activity type which indicates the activity monitor is of limited value in patients with hip osteoarthritis and that the results do not express the real energy expenditure.
Figures


Similar articles
-
Validity of SenseWear® Armband v5.2 and v2.2 for estimating energy expenditure.J Sports Sci. 2016 Oct;34(19):1830-8. doi: 10.1080/02640414.2016.1140220. Epub 2016 Feb 8. J Sports Sci. 2016. PMID: 26854829 Free PMC article.
-
Evaluation of the SenseWear Pro Armband to assess energy expenditure during exercise.Med Sci Sports Exerc. 2004 May;36(5):897-904. doi: 10.1249/01.mss.0000126805.32659.43. Med Sci Sports Exerc. 2004. PMID: 15126727
-
Validation and comparison of two methods to assess human energy expenditure during free-living activities.PLoS One. 2014 Feb 28;9(2):e90606. doi: 10.1371/journal.pone.0090606. eCollection 2014. PLoS One. 2014. PMID: 24587401 Free PMC article.
-
Validation of physical activity monitors in individuals with diabetes: energy expenditure estimation by the multisensor SenseWear Armband Pro3 and the step counter Omron HJ-720 against indirect calorimetry during walking.Diabetes Technol Ther. 2013 May;15(5):413-8. doi: 10.1089/dia.2012.0235. Epub 2013 Apr 4. Diabetes Technol Ther. 2013. PMID: 23557200 Free PMC article.
-
Biomechanical changes and recovery of gait function after total hip arthroplasty for osteoarthritis: a systematic review and meta-analysis.Osteoarthritis Cartilage. 2018 Jul;26(7):847-863. doi: 10.1016/j.joca.2018.02.897. Epub 2018 Feb 21. Osteoarthritis Cartilage. 2018. PMID: 29474993
Cited by
-
Current Physical Activity Monitors in Hip and Knee Osteoarthritis: A Review.Arthritis Care Res (Hoboken). 2017 Oct;69(10):1460-1466. doi: 10.1002/acr.23170. Epub 2017 Sep 6. Arthritis Care Res (Hoboken). 2017. PMID: 27998033 Free PMC article. Review. No abstract available.
-
Validity of physical activity measures in individuals after total knee arthroplasty.Arch Phys Med Rehabil. 2015 Mar;96(3):524-31. doi: 10.1016/j.apmr.2014.10.009. Epub 2014 Oct 31. Arch Phys Med Rehabil. 2015. PMID: 25450127 Free PMC article.
-
Comparison of free-living physical activity measurements between ActiGraph GT3X-BT and Fitbit Charge 3 in young people with haemophilia.Haemophilia. 2022 Nov;28(6):e172-e180. doi: 10.1111/hae.14624. Epub 2022 Jul 13. Haemophilia. 2022. PMID: 35830613 Free PMC article.
-
Active transportation and public transportation use to achieve physical activity recommendations? A combined GPS, accelerometer, and mobility survey study.Int J Behav Nutr Phys Act. 2014 Sep 27;11:124. doi: 10.1186/s12966-014-0124-x. Int J Behav Nutr Phys Act. 2014. PMID: 25260793 Free PMC article.
-
Validation of Activity Tracking Procedures in Elderly Patients after Operative Treatment of Proximal Femur Fractures.Rehabil Res Pract. 2018 Jun 19;2018:3521271. doi: 10.1155/2018/3521271. eCollection 2018. Rehabil Res Pract. 2018. PMID: 30018824 Free PMC article.
References
-
- Troiano RP, Berrigan D, Dodd KW, Mâsse LC, Tilert T, McDowell M. Physical activity in the United States measured by accelerometer. Med Sci Sports Exerc. 2008;40:181–188. - PubMed
-
- Holsgaard-Larsen A, Roos E. Objectively measured physical activity in patients with end stage knee or hip osteoarthritis. Eur J Phys Rehabil Med. 2012;48:1–9. - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources