Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2014 Apr;39(4):1062-8.
doi: 10.1016/j.gaitpost.2014.01.008. Epub 2014 Jan 22.

Accuracy of the Microsoft Kinect sensor for measuring movement in people with Parkinson's disease

Affiliations
Free article

Accuracy of the Microsoft Kinect sensor for measuring movement in people with Parkinson's disease

Brook Galna et al. Gait Posture. 2014 Apr.
Free article

Abstract

Background: The Microsoft Kinect sensor (Kinect) is potentially a low-cost solution for clinical and home-based assessment of movement symptoms in people with Parkinson's disease (PD). The purpose of this study was to establish the accuracy of the Kinect in measuring clinically relevant movements in people with PD.

Methods: Nine people with PD and 10 controls performed a series of movements which were measured concurrently with a Vicon three-dimensional motion analysis system (gold-standard) and the Kinect. The movements included quiet standing, multidirectional reaching and stepping and walking on the spot, and the following items from the Unified Parkinson's Disease Rating Scale: hand clasping, finger tapping, foot, leg agility, chair rising and hand pronation. Outcomes included mean timing and range of motion across movement repetitions.

Results: The Kinect measured timing of movement repetitions very accurately (low bias, 95% limits of agreement <10% of the group mean, ICCs >0.9 and Pearson's r>0.9). However, the Kinect had varied success measuring spatial characteristics, ranging from excellent for gross movements such as sit-to-stand (ICC=.989) to very poor for fine movement such as hand clasping (ICC=.012). Despite this, results from the Kinect related strongly to those obtained with the Vicon system (Pearson's r>0.8) for most movements.

Conclusions: The Kinect can accurately measure timing and gross spatial characteristics of clinically relevant movements but not with the same spatial accuracy for smaller movements, such as hand clasping.

Keywords: Accuracy; Microsoft Kinect; Parkinson's disease; Validity.

PubMed Disclaimer

Publication types