Cross-sectional changes in lung volume measured by electrical impedance tomography are representative for the whole lung in ventilated preterm infants
- PMID: 24561568
- DOI: 10.1097/CCM.0000000000000230
Cross-sectional changes in lung volume measured by electrical impedance tomography are representative for the whole lung in ventilated preterm infants
Abstract
Objective: Electrical impedance tomography measures lung volume in a cross-sectional slice of the lung. Whether these cross-sectional volume changes are representative of the whole lung has only been investigated in adults, showing conflicting results. This study aimed to compare cross-sectional and whole lung volume changes using electrical impedance tomography and respiratory inductive plethysmography.
Design: A prospective, single-center, observational, nonrandomized study.
Setting: The study was conducted in a neonatal ICU in the Netherlands.
Patients: High-frequency ventilated preterm infants with respiratory distress syndrome.
Interventions: Cross-sectional and whole lung volume changes were continuously and simultaneously measured by, respectively, electrical impedance tomography and respiratory inductive plethysmography during a stepwise recruitment procedure. End-expiratory lung volume changes were assessed by mapping the inflation and deflation limbs using both the pressure/impedance and pressure/inductance pairs and characterized by calculating the inflection points. In addition, oscillatory tidal volume changes were assessed at each pressure step.
Measurements and main results: Twenty-three infants were included in the study. Of these, eight infants had to be excluded because the quality of the registration was insufficient for analysis (two electrical impedance tomography and six respiratory inductive plethysmography). In the remaining 15 infants (gestational age 28.0 ± 2.6 wk; birth weight 1,027 ± 514 g), end-expiratory lung volume changes measured by electrical impedance tomography were significantly correlated to respiratory inductive plethysmography measurements in 12 patients (mean r = 0.93 ± 0.05). This was also true for the upper inflection point on the inflation (r = 0.91, p < 0.01) and deflation limb (r = 0.83, p < 0.01). In 13 patients, impedance and inductance data also correlated significantly on oscillatory tidal volume/pressure relationships (mean r = 0.81 ± 0.18).
Conclusions: This study shows that cross-sectional lung volume changes measured by electrical impedance tomography are representative for the whole lung and that this concept also applies to newborn infants.
Similar articles
-
Determination of Lung Volume and Hemodynamic Changes During High-Frequency Ventilation Recruitment in Preterm Neonates With Respiratory Distress Syndrome.Crit Care Med. 2015 Aug;43(8):1685-91. doi: 10.1097/CCM.0000000000000967. Crit Care Med. 2015. PMID: 25803651
-
Changes in lung volume and ventilation during lung recruitment in high-frequency ventilated preterm infants with respiratory distress syndrome.J Pediatr. 2011 Aug;159(2):199-205.e2. doi: 10.1016/j.jpeds.2011.01.066. Epub 2011 Mar 17. J Pediatr. 2011. PMID: 21414632
-
Changes in lung volume and ventilation during surfactant treatment in ventilated preterm infants.Am J Respir Crit Care Med. 2011 Jul 1;184(1):100-5. doi: 10.1164/rccm.201103-0375OC. Epub 2011 Apr 14. Am J Respir Crit Care Med. 2011. PMID: 21493733
-
Noninvasive assessment of lung volume: respiratory inductance plethysmography and electrical impedance tomography.Crit Care Med. 2005 Mar;33(3 Suppl):S163-9. doi: 10.1097/01.ccm.0000155917.39056.97. Crit Care Med. 2005. PMID: 15753723 Review.
-
Bench-to-bedside review: Ventilator strategies to reduce lung injury -- lessons from pediatric and neonatal intensive care.Crit Care. 2005 Apr;9(2):177-83. doi: 10.1186/cc2987. Epub 2004 Nov 4. Crit Care. 2005. PMID: 15774075 Free PMC article. Review.
Cited by
-
Chest electrical impedance tomography examination, data analysis, terminology, clinical use and recommendations: consensus statement of the TRanslational EIT developmeNt stuDy group.Thorax. 2017 Jan;72(1):83-93. doi: 10.1136/thoraxjnl-2016-208357. Epub 2016 Sep 5. Thorax. 2017. PMID: 27596161 Free PMC article. Review.
-
High-flow oxygen therapy versus noninvasive ventilation: a randomised physiological crossover study of alveolar recruitment in acute respiratory failure.ERJ Open Res. 2021 Dec 13;7(4):00373-2021. doi: 10.1183/23120541.00373-2021. eCollection 2021 Oct. ERJ Open Res. 2021. PMID: 34912882 Free PMC article.
-
Early prediction of pulmonary outcomes in preterm infants using electrical impedance tomography.Front Pediatr. 2023 May 24;11:1167077. doi: 10.3389/fped.2023.1167077. eCollection 2023. Front Pediatr. 2023. PMID: 37292377 Free PMC article.
-
Prone positioning improves ventilation-perfusion matching assessed by electrical impedance tomography in patients with ARDS: a prospective physiological study.Crit Care. 2022 May 27;26(1):154. doi: 10.1186/s13054-022-04021-0. Crit Care. 2022. PMID: 35624489 Free PMC article.
-
Lung Recruitment Strategies During High Frequency Oscillatory Ventilation in Preterm Lambs.Front Pediatr. 2019 Jan 22;6:436. doi: 10.3389/fped.2018.00436. eCollection 2018. Front Pediatr. 2019. PMID: 30723711 Free PMC article.
Publication types
MeSH terms
Supplementary concepts
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical