Optimizing evaluation of split renal function in a living kidney donor using scintigraphy and calculation of the geometric mean: a case report

Affiliations
Case Reports

Optimizing evaluation of split renal function in a living kidney donor using scintigraphy and calculation of the geometric mean: a case report

S Weinberger et al. Case Rep Nephrol Urol. .

Abstract

Within the evaluation process of living kidney donors, split renal function is usually evaluated by renal scintigraphy. Since split renal function measured by conventional posterior scans depends on the position of the kidney, actual suitable donors may be rejected because of an inaccurate examination technique. We report the case of a 28-year-old male living kidney donor. Due to a complex vascular anatomy of the right kidney, only his left kidney was considered eligible for transplantation. In conventional posterior Tc99m-mercapto-acetyltriglycine scintigraphy, the left kidney had a relative function of 60%. A second scintigraphy using anterior and posterior dimercaptosuccinic acid scans with calculation of the geometric mean showed an adapted relative function of the left kidney of 53%, now meeting the inclusion criteria for living kidney donation. This case shows that the geometric mean method using simultaneous anterior and posterior views obtained with a dual-head gamma camera can be a very helpful approach to determine split renal function of potential living kidney donors. Further investigation is necessary to prove the benefit of a general bilateral scan before living kidney donation.

Keywords: Geometric mean; Living kidney donor; Renal function; Scintigraphy.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
CT angiogram of the right kidney with complex vessel anatomy.
Fig. 2
Fig. 2
Conventional posterior Tc99m-MAG3 scintigraphy.
Fig. 3
Fig. 3
Posterior and anterior DMSA scintigraphy.

References

    1. Delmonico FL, Council of the Transplantation Society A report of the Amsterdam Forum on the care of the live kidney donor: data and medical guidelines. Transplantation. 2005;79(suppl 6):S53–S66. - PubMed
    1. Ethics Committee of the Transplantation Society The consensus statement of the Amsterdam Forum on the care of the live kidney donor. Transplantation. 2004;78:491–492. - PubMed
    1. Goh YS, Wu MW, Tai BC, Lee KC, Raman L, Teo BW, Vathsala A, Tiong HY. Comparison of creatinine based and kidney volume based methods of estimating glomerular filtration rates in potential living kidney donors. J Urol. 2013;190:1820–1826. - PubMed
    1. Wujanto MB, Lawson RS, Prescott MC, Testa HJ. The importance of using anterior and posterior views in the calculation of differential renal function using 99mTc-DMSA. Br J Radio. 1987;160:869–872. - PubMed
    1. Fleming JS. A technique for analysis of geometric mean renography. Nucl Med Commun. 2006;27:701–708. - PubMed