Observational Study
. 2014 Jan-Feb;8(1):58-66.
doi: 10.1016/j.jcct.2013.12.007. Epub 2014 Jan 11.

Assessment of isotropic calcium using 0.5-mm reconstructions from 320-row CT data sets identifies more patients with non-zero Agatston score and more subclinical atherosclerosis than standard 3.0-mm coronary artery calcium scan and CT angiography

Affiliations
Observational Study

Assessment of isotropic calcium using 0.5-mm reconstructions from 320-row CT data sets identifies more patients with non-zero Agatston score and more subclinical atherosclerosis than standard 3.0-mm coronary artery calcium scan and CT angiography

Anum Aslam et al. J Cardiovasc Comput Tomogr. 2014 Jan-Feb.

Abstract

Background: The presence of calcified plaque in coronary arteries can be quantified by using 0.5-mm isotropic reconstructions from 320-row CT without increased radiation dose. Little is known about reclassification of patients with non-zero Agatston scores and quantitative measures of calcified plaque using 0.5-mm reconstructions.

Objective: The aim was to compare proportions of zero vs. non-zero Agatston scores (subclinical atherosclerosis) in 0.5-mm isotropic reconstructions vs. standard 3.0-mm and CT angiography (CTA) scans on 320-row CT.

Methods: Prospectively, we quantified calcified plaque in coronary arteries in 104 patients by using non-contrast-enhanced scans with 0.5 and 3.0 mm. Coronary calcium assessment was determined by 2 observers. Clinically indicated CTA was also performed; coronary calcium assessment findings were compared with CTA. Ranked Wilcoxon test and χ2 test were performed for comparison. Reproducibility for proportion of zero vs non-zero was assessed by κ statistics.

Results: Median Agatston score (41.9 [interquartile range (IQR), 3.7-213.6] vs. 5.2 [IQR, 0.0-128.5]), calcium volume (53.6 mm3 [IQR, 8.1-202.3] vs. 5.1 mm(3) [IQR, 0.0-96.8],), and lesion number (10.0 [IQR, 3.5-18.5] vs. 1.0 [IQR, 0.0-6.0]) were significantly higher on 0.5-mm reconstruction (P < .0001) than on 3.0-mm reconstruction. More patients with subclinical atherosclerosis were detected on 0.5 mm than on 3.0 mm and CTA scans (76.9% vs. 53.8% vs. 54.8%; P < .0001). The κ values for inter-rater agreement were 0.94 and 0.52 on 3.0- and 0.5-mm data sets, respectively. However, when Agatston scores < 10 were excluded from analysis, the κ value rose to 0.83.

Conclusion: Isotropic 0.5-mm reconstruction detected 23.1% and 22.1% more patients with subclinical atherosclerosis than standard 3.0-mm scans and CTA, which may be more sensitive for the detection of subclinical atherosclerosis; its potential clinical utility needs to be validated in large, prospective studies.

Keywords: 320-Mμltidetector computed tomography; Agatston score; CT angiography; Coronary artery calcium; Image noise; Subclinical atherosclerosis.