Disagreement of the two oscillometric blood pressure measurement devices, Datascope Accutorr Plus and Omron HEM-705CP II, and bidirectional conversion of blood pressure values
- PMID: 24583967
- DOI: 10.1097/MBP.0000000000000037
Disagreement of the two oscillometric blood pressure measurement devices, Datascope Accutorr Plus and Omron HEM-705CP II, and bidirectional conversion of blood pressure values
Abstract
Objective(s): Oscillometric blood pressure (BP) measurement devices frequently replace the standard mercury sphygmomanometer. Comparisons of oscillometric devices are rare, but their agreement is important to ensure comparability of BP data. This study aims to compare two oscillometric devices, Datascope Accutorr Plus and Omron HEM-705CP II, and to develop BP conversion models.
Methods: A sample of 109 adults aged 21-64 years were subjected to alternate same-arm BP measurements according to the International Protocol revision 2010 for the validation of BP measuring devices in adults of the European Society of Hypertension.
Results: A total of 327 BP measurement pairs were obtained. Datascope systolic blood pressure (SBP) pairs, in mmHg, were optimal (<120) for n=188, prehypertensive (120-139) for n=107, and hypertensive (≥140) for n=32 [diastolic blood pressure (DBP)<80 (n=261)/80-89 (n=57)/≥90 (n=9)]. The mean Omron values were higher and the difference increased with BP [mean differences, Omron minus Datascope, within BP ranges were (in mmHg): SBP 1.1±4.7, 3.0±5.5, and 9.3±6.7 and DBP 0.2±3.3, 2.3±3.4, and 5.1±3.9] and pulse pressure (>50 mmHg, SBP difference 5.6±6.3). The prevalence of hypertensive BP was 11% with Omron and 5% with Datascope. Bidirectional conversion models of SBP and DBP values include BP, pulse pressure, age, sex, and the difference in the ratio of cuff width to arm circumference.
Conclusion: The disagreement in oscillometric devices can reach a magnitude that could be of interest for clinical and epidemiological contexts. Conversion formulas with BP, pulse pressure, sex, age, and the cuff width to arm circumference ratio may help to improve comparability.
Similar articles
-
Comparison of two blood pressure oscillometric devices: Datascope Accutorr Plus and Mobil-O-Graph PWA and conversion of blood pressure values from one device to the other.Blood Press Monit. 2020 Feb;25(1):42-49. doi: 10.1097/MBP.0000000000000409. Blood Press Monit. 2020. PMID: 31633519
-
Calibration of blood pressure data after replacement of the standard mercury sphygmomanometer by an oscillometric device and concurrent change of cuffs.Blood Press Monit. 2015 Feb;20(1):39-42. doi: 10.1097/MBP.0000000000000081. Blood Press Monit. 2015. PMID: 25144600
-
Validation of the Omron M5-I, R5-I and HEM-907 automated blood pressure monitors in elderly individuals according to the International Protocol of the European Society of Hypertension.Blood Press Monit. 2007 Aug;12(4):233-42. doi: 10.1097/MBP.0b013e32813fa386. Blood Press Monit. 2007. PMID: 17625396
-
Can an automatic oscillometric device replace a mercury sphygmomanometer on blood pressure measurement? a systematic review and meta-analysis.Blood Press Monit. 2019 Dec;24(6):265-276. doi: 10.1097/MBP.0000000000000412. Blood Press Monit. 2019. PMID: 31658107
-
Accuracy of automated blood pressure measurements in the presence of atrial fibrillation: systematic review and meta-analysis.J Hum Hypertens. 2019 May;33(5):352-364. doi: 10.1038/s41371-018-0153-z. Epub 2019 Jan 10. J Hum Hypertens. 2019. PMID: 30631126
Cited by
-
Hypertension in Germany.Dtsch Arztebl Int. 2016 Dec 2;113(48):809-815. doi: 10.3238/arztebl.2016.0809. Dtsch Arztebl Int. 2016. PMID: 28073425 Free PMC article.
-
Development and Validation of a Hypertension Prevalence Estimator Tool For Use in Clinical Settings.J Clin Hypertens (Greenwich). 2016 Aug;18(8):750-61. doi: 10.1111/jch.12746. Epub 2016 Jan 5. J Clin Hypertens (Greenwich). 2016. PMID: 26729615 Free PMC article.
-
Single-Pill, Triple Antihypertensive Therapy in Rural Sub-Saharan Africa: Preliminary Experience.Cardiol Ther. 2024 Jun;13(2):431-442. doi: 10.1007/s40119-024-00358-5. Epub 2024 Feb 12. Cardiol Ther. 2024. PMID: 38345713 Free PMC article.
-
[Blood pressure measurement in the NAKO German National Cohort (GNC) - differences in methods, distribution of blood pressure values, and awareness of hypertension compared to other population-based studies in Germany].Bundesgesundheitsblatt Gesundheitsforschung Gesundheitsschutz. 2020 Apr;63(4):452-464. doi: 10.1007/s00103-020-03109-8. Bundesgesundheitsblatt Gesundheitsforschung Gesundheitsschutz. 2020. PMID: 32170398 German. No abstract available.
-
Impact of Cardiovascular Risk Factors on Arterial Stiffness in a Countryside Area of Switzerland: Insights from the Swiss Longitudinal Cohort Study.Cardiol Ther. 2022 Dec;11(4):545-557. doi: 10.1007/s40119-022-00280-8. Epub 2022 Sep 24. Cardiol Ther. 2022. PMID: 36152116 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical