Efficacy of high-intensity focused ultrasonography for noninvasive body sculpting in Chinese patients
- PMID: 24585511
- DOI: 10.1002/lsm.22232
Efficacy of high-intensity focused ultrasonography for noninvasive body sculpting in Chinese patients
Abstract
Background and objectives: High-intensity focused ultrasound (HIFU) is a noninvasive technology for body contouring. HIFU is focused within the subcutaneous adipose tissue, causing coagulative necrosis and cell death. The objective of this study is to evaluate the effectiveness of a HIFU device for sculpting of the abdomen.
Materials and methods: The system has a set focal depth of 1.3 cm. Twelve subjects with adipose thickness no less than 2.5 cm who met the screening criteria were recruited. Each subject received one treatment to the abdomen. The total fluence used per site was 150-165 J/cm(2) with a mean of 161 J/cm(2) . The waist circumference at iliac crest and the point of maximum circumference were recorded at baseline, 4, 8, and 12 weeks post-treatment, as well as their weight and BMI. Subjects' rating on comfort level and satisfaction were collected via questionnaires at every follow-up. Standardized photographs were also taken with the Canfield System® at each visit.
Results: Seven out of 12 subjects were satisfied with the outcome and 9 out of 12 would recommend this treatment to their friends and family. There was statistically significant improvement in the waist circumference measured at both the iliac crest (P-value 0.013, 0.002, 0.005) and maximum waistline (P-value 0.003, 0.034, 0.023) at 4, 8, 12 weeks post-treatment. Spearman's rho for correlation of energy level versus improvement showed that at 12 weeks post-treatment follow-up, the improvement significantly correlated with the total fluence per treatment (P-value 0.041). The higher the total fluence delivered, the larger the decrease in waist circumference.
Conclusions: High-intensity focused ultrasound effectively decreases waist circumference in Chinese. The higher the total fluence delivered, the larger the decrease in waist circumference was observed.
Keywords: Chinese; body sculpting; high-intensity focused ultrasound; non-invasive.
© 2014 Wiley Periodicals, Inc.
Similar articles
-
Randomized sham-controlled trial to evaluate the safety and effectiveness of a high-intensity focused ultrasound device for noninvasive body sculpting.Plast Reconstr Surg. 2011 Jul;128(1):253-262. doi: 10.1097/PRS.0b013e3182174278. Plast Reconstr Surg. 2011. PMID: 21701341 Clinical Trial.
-
Safety and tolerability of high-intensity focused ultrasonography for noninvasive body sculpting: 24-week data from a randomized, sham-controlled study.Aesthet Surg J. 2012 Sep;32(7):868-76. doi: 10.1177/1090820X12455190. Aesthet Surg J. 2012. PMID: 22942114 Clinical Trial.
-
High-intensity focused ultrasound for the reduction of subcutaneous adipose tissue using multiple treatment techniques.Dermatol Surg. 2014 Jun;40(6):641-51. doi: 10.1111/dsu.0000000000000022. Dermatol Surg. 2014. PMID: 24852468 Clinical Trial.
-
Noninvasive body sculpting technologies with an emphasis on high-intensity focused ultrasound.Aesthetic Plast Surg. 2011 Oct;35(5):901-12. doi: 10.1007/s00266-011-9700-5. Epub 2011 Apr 1. Aesthetic Plast Surg. 2011. PMID: 21461627 Review.
-
Effectiveness of HIFU Therapy for Nonsurgical Facial and Body Contouring: A Systematic Review of Prospective and Experimental Studies.Plast Reconstr Surg. 2023 Mar 1;151(3):533-544. doi: 10.1097/PRS.0000000000009912. Epub 2022 Nov 21. Plast Reconstr Surg. 2023. PMID: 36730550
Cited by
-
Cerium-doped calcium carbonate microparticles combined with low-intensity ultrasound for efficient sonodynamic therapy in body sculpting.J Biol Eng. 2025 Apr 28;19(1):39. doi: 10.1186/s13036-025-00505-z. J Biol Eng. 2025. PMID: 40296005 Free PMC article.
-
Efficacy and safety of a novel combined 1060-nm and 635-nm laser device for non-invasive reduction of abdominal and submental fat.Lasers Med Sci. 2022 Feb;37(1):505-512. doi: 10.1007/s10103-021-03288-z. Epub 2021 Apr 2. Lasers Med Sci. 2022. PMID: 33797649 Clinical Trial.
-
Hepathic, biochemical, hematological, and histological effects of the ultracavitation in rabbits livers.Acta Cir Bras. 2020 Jun 19;35(4):e202000403. doi: 10.1590/s0102-865020200040000003. eCollection 2020. Acta Cir Bras. 2020. PMID: 32578723 Free PMC article.
-
Use of Stationary Ultrasound Device In Subcutaneous Abdominal Fat Tissue: Analysis of Clinical Effects.J Clin Aesthet Dermatol. 2023 Feb;16(2):19-23. J Clin Aesthet Dermatol. 2023. PMID: 36909872 Free PMC article.
-
[Microfocussed vs. macrofocussed ultrasound for body contouring].Dermatologie (Heidelb). 2023 Oct;74(10):753-758. doi: 10.1007/s00105-023-05199-2. Epub 2023 Aug 14. Dermatologie (Heidelb). 2023. PMID: 37578555 Review. German.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous