. 2014 Feb;6(1):35-8.
doi: 10.4047/jap.2014.6.1.35. Epub 2014 Feb 14.

Correlation between microleakage and screw loosening at implant-abutment connection

Affiliations

Correlation between microleakage and screw loosening at implant-abutment connection

Cem Sahin et al. J Adv Prosthodont. 2014 Feb.

Abstract

Purpose: This study aimed to evaluate the correlation between microleakage and screw loosening at different types of implant-abutment connections and/or geometries measuring the torque values before and after the leakage tests.

Materials and methods: Three different abutment types (Intenal hex titanium, internal hex zirconium, morse tapered titaniuım) with different geometries were connected to its own implant fixture. All the abutments were tightened with a standard torque value then the composition was connected to the modified fluid filtration system. After the measurements of leakage removal torque values were re-measured. Kruskal-wallis test was performed for non-parametric and one-way ANOVA was performed for parametric data. The correlation was evaluated using Spearman Correlation Test (α=0.05).

Results: Significantly higher microleakage was found at the connection of implant-internal hex zirconium abutment. Observed mean torque value loss was also significantly higher than other connection geometries. Spearman tests revealed a significant correlation between microleakage and screw loosening.

Conclusion: Microleakage may provoke screw loosening. Removing torque values rationally decrease with the increase of microleakage.

Keywords: Implant-abutment connection; Microleakage; Screw loosening.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
Schematic view of microleakage throuh the screw threads and implant-abutment interface.

References

    1. McGlumphy EA, Mendel DA, Holloway JA. Implant screw mechanics. Dent Clin North Am. 1998;42:71–89. - PubMed
    1. Byrne D, Jacobs S, O'Connell B, Houston F, Claffey N. Preloads generated with repeated tightening in three types of screws used in dental implant assemblies. J Prosthodont. 2006;15:164–171. - PubMed
    1. Binon PP. Implants and components: entering the new millennium. Int J Oral Maxillofac Implants. 2000;15:76–94. - PubMed
    1. Pjetursson BE, Brägger U, Lang NP, Zwahlen M. Comparison of survival and complication rates of tooth-supported fixed dental prostheses (FDPs) and implant-supported FDPs and single crowns (SCs) Clin Oral Implants Res. 2007;18:97–113. - PubMed
    1. Jung RE, Pjetursson BE, Glauser R, Zembic A, Zwahlen M, Lang NP. A systematic review of the 5-year survival and complication rates of implant-supported single crowns. Clin Oral Implants Res. 2008;19:119–130. - PubMed

LinkOut - more resources