Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2014 Mar 8;13(1):24.
doi: 10.1186/1475-925X-13-24.

Reconstruction of dual-frequency conductivity by optimization of phase map in MREIT and MREPT

Affiliations

Reconstruction of dual-frequency conductivity by optimization of phase map in MREIT and MREPT

Oh In Kwon et al. Biomed Eng Online. .

Abstract

Background: The spectroscopic conductivity distribution of tissue can help to explain physiological and pathological status. Dual frequency conductivity imaging by combining Magnetic Resonance Electrical Property Tomography (MREPT) and Magnetic Resonance Electrical Impedance Tomography (MREIT) has been recently proposed. MREIT can provide internal conductivity distributions at low frequency (below 1 kHz) induced by an external injecting current. While MREPT can provide conductivity at the Larmor frequency related to the strength of the magnetic field. Despite this potential to describe the membrane properties using spectral information, MREPT and MREIT techniques currently suffer from weak signals and noise amplification as they both reply on differentiation of measured phase data.

Methods: We proposed a method to optimize the measured phase signal by finding weighting factors according to the echo signal for MREPT and MREIT using the ICNE (Injected current nonlinear encoding) multi-echo pulse sequence. Our target weights are chosen to minimize the measured noise. The noise standard deviations were precisely analyzed for the optimally weighted magnetic flux density and the phase term of the positive-rotating magnetic field. To enhance the quality of dual-frequency conductivity images, we applied the denoising method based on the reaction-diffusion equation with the estimated noise standard deviations. A real experiment was performed with a hollow cylindrical object made of thin insulating film with holes to control the apparent conductivity using ion mobility and an agarose gel cylinder wrapped in an insulating film without holes to show different spectroscopic conductivities.

Results: The ability to image different conductivity characteristics in MREPT and MREIT from a single MR scan was shown by including the two objects with different spectroscopic conductivities. Using the six echo signals, we computed the optimized weighting factors for each echo. The qualities of conductivity images for MREPT and MREIT were improved by optimization of the phase map. The proposed method effectively reduced the random noise artifacts for both MREIT and MREPT.

Conclusion: We enhanced the dual conductivity images using the optimally weighted magnetic flux density and the phase term of positive-rotating magnetic field based on the analysis of the noise standard deviations and applying the optimization and denoising methods.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Schematic diagram for ICNE multi-echo pulse sequence to acquire dual-frequency conductivity images. ‘RF’ shows the timing for excitation of 90° and 180° RF pulses. Gz, Gy and Gx are slice selection gradient, phase encoding gradient and readout gradient, respectively. ζj± is the j-th echo signal. ‘ICNE’ shows the alternating injection currents in the form of pulses synchronized with the RF pulse.
Figure 2
Figure 2
Experimental set-up for a phantom.(a) The phantom configuration at the middle slice of the testing object, (b) a cylindrical phantom with saline solution of 0.2 Sm−1 including a thin film object and an agarose anomaly wrapped by an insulating film, and (c) a magnitude image at the middle slice acquired at TE=15 ms.
Figure 3
Figure 3
MR magnitude images measured at each echo time. Echo number marked on the upper left part of each image as (1), (2), ⋯, (6). Magnitude images using ICNE multi-echo pulse sequence acquired at the echo time TEj=15×j ms for j=1,⋯,6.
Figure 4
Figure 4
Phase images and magnetic flux density images of each echo.(a) Phase images φj+ of H+ map in radian and (b) magnetic flux density images Bzj in ’nT’ unit using ICNE multi-echo pulse sequence at the echo time TEj=15×j ms for j=1,⋯,6.
Figure 5
Figure 5
Images of weighting factors corresponding to each echo.(a) Images of weighting factors corresponding to each echoes for φj+ and (b) images for magnetic flux density Bzj to produce an optimal φ+,χ and Bzξ, respectively. The weighting factors were scaled from 0 to 1.
Figure 6
Figure 6
Reconstructed conductivity images of each echo.(a) Reconstructed conductivity images using the multiple echoes φj+ for j=1,⋯,6 at Larmor frequency and (b) conductivity images using Bzjd for j=1,⋯,6 at the low frequency.
Figure 7
Figure 7
Optimized dual-frequency conductivity images with and without applying the optimization and denoising method.(a) and (e) Normalized noise standard deviation of the optimally weighted phase of H+andBzξ, respectively. (b) and (f) Reconstructed conductivity images without using the optimization and denoising method. (c) and (d) Reconstructed conductivity distributions at Larmor frequency using the optimized φ+,χwithout denoising and with the proposed denoising technique applying to ∇φ+,χ, respectively. (g) and (h) Reconstructed low-frequency conductivity distributions without denoising and with the proposed denoising technique applying toBzd,ξ,d=1,2, respectively.

References

    1. Geddes LA, Baker LE. The specific resistance of biological material—a compendium of data for the biomedical engineer and physiologist. Med Biol Eng. 1967;5(3):271–293. doi: 10.1007/BF02474537. - DOI - PubMed
    1. Gabriel S, Lau RW, Gabriel C. The dielectric properties of biological tissues: Ii. measurements in the frequency range 10 hz to 20 ghz. Phys Med Biol. 1996;41(11):2251–2269. doi: 10.1088/0031-9155/41/11/002. - DOI - PubMed
    1. Holder DS. Electrical Impedance Tomography: Methods, History and Applications. Bristol, UK: IOP Publishing; 2005.
    1. Ider YZ, Birgul O. Use of the magnetic field generated by the internal distribution of injected currents for electrical impedance tomography (mr-eit) Elektrik. 1998;6(3):215–225.
    1. Kwon O, Woo EJ, Yoon J-R, Seo JK. Magnetic resonance electrical impedance tomography (mreit): simulation study of j-substitution algorithm. Biomed Eng, IEEE Trans. 2002;49(2):160–167. doi: 10.1109/10.979355. - DOI - PubMed

Publication types