Accuracy of a simplified equation for energy expenditure based on bedside volumetric carbon dioxide elimination measurement--a two-center study
- PMID: 24636151
- PMCID: PMC4670754
- DOI: 10.1016/j.clnu.2014.02.008
Accuracy of a simplified equation for energy expenditure based on bedside volumetric carbon dioxide elimination measurement--a two-center study
Abstract
Background & aims: Accurate assessment of resting energy expenditure (REE) and metabolic state is essential to optimize nutrient intake in critically ill patients. We aimed to examine the accuracy of a simplified equation for predicting REE using carbon dioxide elimination (VCO2) values.
Methods: We conducted a two-center study of metabolic data from mechanically ventilated children less than 18 years of age. Mean respiratory quotient (RQ) from the derivation set (n = 72 subjects) was used to modify the Weir equation to obtain a simplified equation based on VCO2 measurements alone. This equation was then applied to subjects at the second institution (validation dataset, n = 94) to predict resting energy expenditure. Bland-Altman analysis was used to assess the agreement between measured REE values, and REE estimated by the new equation as well as the Schofield equation. We also examined the accuracy of the new equation in classifying patients according to their metabolic state.
Results: Mean respiratory quotient (± SD) of 0.89 ± 0.09 in the derivation set was used to obtain a simplified equation, REE (kcal/day) = 5.534*VCO2 (L/min)*1440. In relation to the measured REE in the validation set, the mean bias (limits of agreement) for the REE predicted by this equation was -0.65% (-14.4-13.1%); and the overall diagnostic accuracy for classifying subjects as hypometabolic or hypermetabolic was 84%. Mean bias (limits) of agreement between measured and Schofield equation estimated REE was -0.1% (-40.5-40.7%).
Conclusions: A simplified metabolic equation using VCO2 values was superior to the standard equation in estimating REE, and provided a reasonably accurate metabolic classification in mechanically ventilated children. In the absence of indirect calorimetry, bedside VCO2 monitoring could provide valuable continuous metabolic information to guide optimal nutrient intake.
Keywords: Energy expenditure; Indirect calorimetry; Metabolic assessment; Respiratory quotient; VCO(2).
Copyright © 2014 Elsevier Ltd and European Society for Clinical Nutrition and Metabolism. All rights reserved.
Conflict of interest statement
References
-
- Brandi LS, Bertolini R, Santini L, Cavani S. Effects of ventilator resetting on indirect calorimetry measurement in the critically ill surgical patient. Critical care medicine. 1999 Mar;27(3):531–9. Epub 1999/04/13. eng. - PubMed
-
- Mehta NM, Compher C. A.S.P.E.N. Clinical Guidelines: nutrition support of the critically ill child. JPEN J Parenter Enteral Nutr. 2009 May-Jun;33(3):260–76. Epub 2009/04/29. eng. - PubMed
-
- Martindale RG, McClave SA, Vanek VW, McCarthy M, Roberts P, Taylor B, et al. Guidelines for the provision and assessment of nutrition support therapy in the adult critically ill patient: Society of Critical Care Medicine and American Society for Parenteral and Enteral Nutrition: Executive Summary. Critical care medicine. 2009 May;37(5):1757–61. Epub 2009/04/18. eng. - PubMed
-
- van der Kuip M, Oosterveld MJ, van Bokhorst-de van der Schueren MA, de Meer K, Lafeber HN, Gemke RJ. Nutritional support in 111 pediatric intensive care units: a European survey. Intensive Care Med. 2004 Sep;30(9):1807–13. eng. - PubMed
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources