Multicenter study of subjective acceptance during magnetic resonance imaging at 7 and 9.4 T
- PMID: 24637589
- DOI: 10.1097/RLI.0000000000000035
Multicenter study of subjective acceptance during magnetic resonance imaging at 7 and 9.4 T
Abstract
Objectives: The aims of this study were to investigate the subjective discomfort and sensory side effects during ultrahigh field (UHF) magnetic resonance imaging (MRI) examinations in a large-scale study and to evaluate differences between magnetic resonance (MR) sites.
Materials and methods: Four MR sites with a 7-T MR system and 2 MR sites with a 9.4-T MR system participated in this multicenter study with a total number of 3457 completed questionnaires on causes of discomfort and sensations during the examination. For a pooled retrospective analysis of the results from the partially different questionnaires, all data were adapted to an answer option with a 4-point scale (0 = no discomfort/side effect, 3 = very unpleasant/very strong sensation). To differentiate effects evoked by the low-frequency time-varying magnetic fields due to movement through the static magnetic field, most questionnaires separated the manifestation of sensory side effects during movement on the patient table from manifestation while lying still in the isocenter.
Results: In general, a high acceptance of UHF examinations was found, where in 82% of the completed questionnaires, the subjects stated the examination to be at least tolerable. Although in 7.6% of the questionnaires, subjects felt discomfort during the examination, only 0.9% of the image acquisitions had to be terminated prematurely. No adverse events occurred in any of the examinations. Only 1% of the subjects were unwilling to undergo further UHF MRI examinations. Examination duration was the most complained cause of discomfort, followed by acoustic noise and lying still. All magnetic-field-related sensations were more pronounced when moving the patient table versus the isocenter position (19%/2% of the subjects felt unpleasant vertigo during the moving/stationary state). In general, vertigo was the most often stated sensory side effect and was more pronounced at 9.4 T compared with 7 T. However, the results varied substantially among the different sites.
Conclusions: The high levels of subjective acceptance found in this study lead to the conclusion that UHF MRI would be tolerated as a diagnostic tool in clinical practice. For more consistent data ascertainment, we propose a standardized questionnaire for subjective perception monitoring.
Similar articles
-
A large-scale study on subjective perception of discomfort during 7 and 1.5 T MRI examinations.Bioelectromagnetics. 2011 Dec;32(8):610-9. doi: 10.1002/bem.20680. Epub 2011 May 19. Bioelectromagnetics. 2011. PMID: 21598286 Clinical Trial.
-
Assessment of sensations experienced by subjects during MR imaging examination at 7T.Magn Reson Med Sci. 2015;14(1):35-41. doi: 10.2463/mrms.2014-0004. Epub 2014 Oct 27. Magn Reson Med Sci. 2015. PMID: 25345414
-
Subject tolerance of 7 T MRI examinations.J Magn Reson Imaging. 2013 Sep;38(3):722-5. doi: 10.1002/jmri.23904. Epub 2012 Nov 13. J Magn Reson Imaging. 2013. PMID: 23150466
-
Static magnetic field effects on human subjects related to magnetic resonance imaging systems.Prog Biophys Mol Biol. 2005 Feb-Apr;87(2-3):255-65. doi: 10.1016/j.pbiomolbio.2004.08.012. Prog Biophys Mol Biol. 2005. PMID: 15556664 Review.
-
Effects of static magnetic fields on cognition, vital signs, and sensory perception: a meta-analysis.J Magn Reson Imaging. 2011 Oct;34(4):758-63. doi: 10.1002/jmri.22720. Epub 2011 Jul 12. J Magn Reson Imaging. 2011. PMID: 21751291 Review.
Cited by
-
Patient Comfort in Modern Computed Tomography: What Really Counts.Tomography. 2022 May 23;8(3):1401-1412. doi: 10.3390/tomography8030113. Tomography. 2022. PMID: 35645399 Free PMC article.
-
Controlled Release of Therapeutics from Thermoresponsive Nanogels: A Thermal Magnetic Resonance Feasibility Study.Cancers (Basel). 2020 May 27;12(6):1380. doi: 10.3390/cancers12061380. Cancers (Basel). 2020. PMID: 32471299 Free PMC article.
-
Decrease of 7T MR short-term effects with repeated exposure.Neuroradiology. 2024 Apr;66(4):567-575. doi: 10.1007/s00234-024-03292-4. Epub 2024 Jan 25. Neuroradiology. 2024. PMID: 38270624 Free PMC article.
-
Subjective perception of safety in healthy individuals working with 7 T MRI scanners: a retrospective multicenter survey.MAGMA. 2016 Jun;29(3):379-87. doi: 10.1007/s10334-016-0527-6. Epub 2016 Feb 9. MAGMA. 2016. PMID: 26861049
-
Induction of Extremely Severe Nausea via Vestibular Activation on a 7 Tesla MRI Scanner.J Magn Reson Imaging. 2025 Jun;61(6):2644-2647. doi: 10.1002/jmri.29656. Epub 2024 Nov 7. J Magn Reson Imaging. 2025. PMID: 39506914 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical