A survey of some pre-analytical errors identified from the Biochemistry Department of a Scottish hospital
- PMID: 24642776
- DOI: 10.1177/0036933014529056
A survey of some pre-analytical errors identified from the Biochemistry Department of a Scottish hospital
Abstract
Background: It is estimated that 70% of all health care decisions affecting diagnosis or treatment involve laboratory testing. The pre-analytical phase of the testing process shows the highest prevalence of errors accounting for 70% of all mistakes in laboratory diagnostics. It is recommended that laboratories collect statistics on pre-analytical error rates. This survey examined some mistakes in blood collection, i.e. specimen haemolysis, and EDTA contamination.
Methods: Survey was from June 2011 to May 2012. Haemolysis was detected by analysers' automated haemolysis index function. Plasma EDTA was measured by an automated system. Data were captured from our laboratory information management system.
Results: For a total workload of 763,577 blood specimens, the overall haemolysis rate was 3.2%. Much higher rates of both specimen haemolysis and EDTA contamination were observed when blood was not collected by trained phlebotomists.
Conclusions: Better training in blood collection, achieving the standard of professional phlebotomists, will improve validity of diagnostic information; reduce risks of dangerous misinterpretation of results, unwanted anaemia and needlestick injury and decrease laboratory supplies costs. This type of audit could be replicated in other Scottish Health Boards with some benefit and thereby better target future training needs.
Keywords: EDTA contamination; Phlebotomy; in vitro haemolysis; training.
Similar articles
-
Pre-analytical workstations: a tool for reducing laboratory errors.Clin Chim Acta. 2009 Jun;404(1):68-74. doi: 10.1016/j.cca.2009.03.024. Epub 2009 Mar 18. Clin Chim Acta. 2009. PMID: 19302988
-
[Pre-analytical errors in blood sampling: an observational study on specimen collection technique and laboratory results].Assist Inferm Ric. 2015 Apr-Jun;34(2):76-85. doi: 10.1702/1942.21102. Assist Inferm Ric. 2015. PMID: 26228504 Italian.
-
Level of confidence in venepuncture and knowledge in determining causes of blood sample haemolysis among clinical staff and phlebotomists.J Clin Nurs. 2015 Feb;24(3-4):370-85. doi: 10.1111/jocn.12607. Epub 2014 May 9. J Clin Nurs. 2015. PMID: 24810794
-
Specimen collection and handling.Southeast Asian J Trop Med Public Health. 1999;30 Suppl 3:158-65. Southeast Asian J Trop Med Public Health. 1999. PMID: 10926276 Review.
-
Haemolysis: an overview of the leading cause of unsuitable specimens in clinical laboratories.Clin Chem Lab Med. 2008;46(6):764-72. doi: 10.1515/CCLM.2008.170. Clin Chem Lab Med. 2008. PMID: 18601596 Review.
Cited by
-
Current Methods of Haemolysis Detection and Reporting as a Source of Risk to Patient Safety: a Narrative Review.Clin Biochem Rev. 2016 Dec;37(4):143-151. Clin Biochem Rev. 2016. PMID: 28167844 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources