The development of a 200 kV monochromated field emission electron source
- PMID: 24657419
- DOI: 10.1016/j.ultramic.2014.02.004
The development of a 200 kV monochromated field emission electron source
Abstract
We report the development of a monochromator for an intermediate-voltage aberration-corrected electron microscope suitable for operation in both STEM and TEM imaging modes. The monochromator consists of two Wien filters with a variable energy selecting slit located between them and is located prior to the accelerator. The second filter cancels the energy dispersion produced by the first filter and after energy selection forms a round monochromated, achromatic probe at the specimen plane. The ultimate achievable energy resolution has been measured as 36 meV at 200 kV and 26 meV at 80 kV. High-resolution Annular Dark Field STEM images recorded using a monochromated probe resolve Si-Si spacings of 135.8 pm using energy spreads of 218 meV at 200 kV and 217 meV at 80 kV respectively. In TEM mode an improvement in non-linear spatial resolution to 64 pm due to the reduction in the effects of partial temporal coherence has been demonstrated using broad beam illumination with an energy spread of 134 meV at 200 kV.
Keywords: Aberration correction; Electron energy-loss spectroscopy; Energy resolution; High-resolution imaging; Monochromator; Wien filter.
Copyright © 2014 Elsevier B.V. All rights reserved.
Similar articles
-
Development of a monochromator for aberration-corrected scanning transmission electron microscopy.Microscopy (Oxf). 2015 Jun;64(3):151-8. doi: 10.1093/jmicro/dfv001. Epub 2015 Feb 5. Microscopy (Oxf). 2015. PMID: 25654985
-
First experimental test of a new monochromated and aberration-corrected 200 kV field-emission scanning transmission electron microscope.Ultramicroscopy. 2006 Oct-Nov;106(11-12):963-9. doi: 10.1016/j.ultramic.2006.04.014. Epub 2006 Jul 5. Ultramicroscopy. 2006. PMID: 16870338
-
Monochromated STEM with a 30 meV-wide, atom-sized electron probe.Microscopy (Oxf). 2013 Feb;62(1):3-21. doi: 10.1093/jmicro/dfs089. Epub 2013 Jan 17. Microscopy (Oxf). 2013. PMID: 23335810
-
Using a monochromator to improve the resolution in TEM to below 0.5Å. Part I: Creating highly coherent monochromated illumination.Ultramicroscopy. 2012 Mar;114:72-81. doi: 10.1016/j.ultramic.2012.01.008. Epub 2012 Jan 25. Ultramicroscopy. 2012. PMID: 22356791
-
Resolving 45-pm-separated Si-Si atomic columns with an aberration-corrected STEM.Microscopy (Oxf). 2015 Jun;64(3):213-7. doi: 10.1093/jmicro/dfv014. Epub 2015 Mar 29. Microscopy (Oxf). 2015. PMID: 25825509
Cited by
-
CryoEM at 100 keV: a demonstration and prospects.IUCrJ. 2019 Oct 11;6(Pt 6):1086-1098. doi: 10.1107/S2052252519012612. eCollection 2019 Nov 1. IUCrJ. 2019. PMID: 31709064 Free PMC article.
-
Exploring the capabilities of monochromated electron energy loss spectroscopy in the infrared regime.Sci Rep. 2018 Apr 4;8(1):5637. doi: 10.1038/s41598-018-23805-5. Sci Rep. 2018. PMID: 29618757 Free PMC article.
-
Snapshot 3D Electron Imaging of Structural Dynamics.Sci Rep. 2017 Sep 7;7(1):10839. doi: 10.1038/s41598-017-10654-x. Sci Rep. 2017. PMID: 28883400 Free PMC article.
-
Practical aspects of monochromators developed for transmission electron microscopy.Microscopy (Oxf). 2014 Oct;63(5):337-44. doi: 10.1093/jmicro/dfu027. Epub 2014 Aug 14. Microscopy (Oxf). 2014. PMID: 25125333 Free PMC article.
LinkOut - more resources
Full Text Sources
Other Literature Sources