. 2014 Apr 2;9(4):e93368.
doi: 10.1371/journal.pone.0093368. eCollection 2014.

Persistence of microbial contamination on transvaginal ultrasound probes despite low-level disinfection procedure

Affiliations

Persistence of microbial contamination on transvaginal ultrasound probes despite low-level disinfection procedure

Fatima M'Zali et al. PLoS One. .

Abstract

Aim of the study: In many countries, Low Level Disinfection (LLD) of covered transvaginal ultrasound probes is recommended between patients' examinations. The aim of this study was to evaluate the antimicrobial efficacy of LLD under routine conditions on a range of microorganisms.

Materials and methods: Samples were taken over a six month period in a private French Radiology Center. 300 specimens derived from endovaginal ultrasound probes were analyzed after disinfection of the probe with wipes impregnated with a quaternary ammonium compound and chlorhexidine. Human papillomavirus (HPV) was sought in the first set of s100 samples, Chlamydia trachomatis and mycoplasmas were searched in the second set of 100 samples, bacteria and fungi in the third 100 set samples. HPV, C. trachomatis and mycoplasmas were detected by PCR amplification. PCR positive samples were subjected to a nuclease treatment before an additional PCR assay to assess the likely viable microorganisms. Bacteria and fungi were investigated by conventional methods.

Results: A substantial persistence of microorganisms was observed on the disinfected probes: HPV DNA was found on 13% of the samples and 7% in nuclease-resistant form. C. trachomatis DNA was detected on 20% of the probes by primary PCR but only 2% after nuclease treatment, while mycoplasma DNA was amplified in 8% and 4%, respectively. Commensal and/or environmental bacterial flora was present on 86% of the probes, occasionally in mixed culture, and at various levels (10->3000 CFU/probe); Staphylococcus aureus was cultured from 4% of the probes (10-560 CFU/probe). No fungi were isolated.

Conclusion: Our findings raise concerns about the efficacy of impregnated towels as a sole mean for disinfection of ultrasound probes. Although the ultrasound probes are used with disposable covers, our results highlight the potential risk of cross contamination between patients during ultrasound examination and emphasize the need for reviewing the disinfection procedure.

PubMed Disclaimer

Conflict of interest statement

Competing Interests: The authors have declared that no competing interests exist.

References

    1. Rutala WA, Weber DJ, and the Healthcare Infection Control Practices Advisory Committee (HICPAC) (2008) Guideline for disinfection and sterilization in healthcare facilities. Available: http://www.cdc.gov/hicpac/pdf/guidelines/disinfection_nov_2008.pdf. Accessed 2013 May 7.
    1. American Institute of Ultrasound in Medicine (2003) Guidelines for Cleaning and Preparing Endocavitary Ultrasound Transducers Between Patients. Ultrasound Obstet Gynecol 7: 94.
    1. Gray RA, Williams PL, Dubbins PA, Jenks PJ (2012) Decontamination of transvaginal ultrasound probes: review of national practice and need for national guidelines. Clin Radiol 67: 1069–77. - PubMed
    1. Backhouse S (2003) Establishing a protocol for the cleaning and sterilisation/disinfection of ultrasound transducers. BMUS Bulletin 11: 37–39.
    1. Haut Conseil de la Santé Publique (2007) Gaines de protection à usage unique pour dispositifs médicaux réutilisables: recommandations d'utilisation. Rapport, décembre 2007. Available: http://www.hcsp.fr/explore.cgi/hcspr20071214_GainesProtec.pdf. Accessed 2013 May 7.