. 2014 Mar 6;15(2):4523.
doi: 10.1120/jacmp.v15i2.4523.

Relative stopping power measurements to aid in the design of anthropomorphic phantoms for proton radiotherapy

Affiliations

Relative stopping power measurements to aid in the design of anthropomorphic phantoms for proton radiotherapy

Ryan L Grant et al. J Appl Clin Med Phys. .

Abstract

The delivery of accurate proton dose for clinical trials requires that the appropriate conversion function from Hounsfield unit (HU) to relative linear stopping power (RLSP) be used in proton treatment planning systems (TPS). One way of verifying that the TPS is calculating the correct dose is an end-to-end test using an anthropomorphic phantom containing tissue equivalent materials and dosimeters. Many of the phantoms in use for such end-to-end tests were originally designed using tissue-equivalent materials that had physical characteristics to match patient tissues when irradiated with megavoltage photon beams. The aim of this study was to measure the RLSP of materials used in the phantoms, as well as alternative materials to enable modifying phantoms for use at proton therapy centers. Samples of materials used and projected for use in the phantoms were measured and compared to the HU assigned by the treatment planning system. A percent difference in RLSP of 5% was used as the cutoff for materials deemed acceptable for use in proton therapy (i.e., proton equivalent). Until proper tissue-substitute materials are identified and incorporated, institutions that conduct end-to-end tests with the phantoms are instructed to override the TPS with the measured stopping powers we provide. To date, the RLSPs of 18 materials have been measured using a water phantom and/or multilayer ion chamber (MLIC). Nine materials were identified as acceptable for use in anthropomorphic phantoms. Some of the failing tissue substitute materials are still used in the current phantoms. Further investigation for additional appropriate tissue substitute materials in proton beams is ongoing. Until all anthropomorphic phantoms are constructed of appropriate materials, a unique HU-RLSP phantom has been developed to be used during site visits to verify the proton facility's treatment planning HU-RLSP calibration curve.

PubMed Disclaimer

Figures

Figure 1
Figure 1
A portion of the depth dose curve in water with (circles) and without (squares) Nylon added to the surface. The curve is a cubic spline fit to the data. “x” indicates the location of the distal 80% dose of each curve.
Figure 2
Figure 2
Materials with measured RLSP and assigned RLSP. Materials circled were included in the HU‐RLSP phantom, shown in Fig. 3. Error bars indicate uncertainty expressed as 1 standard deviation in the measured value.
Figure 3
Figure 3
HU‐RLSP phantom and CT image of the phantom.

References

    1. Ibbott GS, Followill DS, Molineu HA, Lowenstein JR, Alvarez PE, Roll JE. Challenges in credentialing institutions and participants in advanced technology multi‐institutional clinical trials. Int J Radiat Oncol Biol Phys. 2008;71(1 Suppl):S71–S75. - PMC - PubMed
    1. Followill DS, Evans DR, Cherry C, et al. Design, development, and implementation of the radiological physics center's pelvis and thorax anthropomorphic quality assurance phantoms. Med Phys. 2007;34(6):2070–76. - PubMed
    1. Molineu A, Followill DS, Balter PA, et al. Design and implementation of an anthropomorphic quality assurance phantom for intensity‐modulated radiation therapy for the Radiation Therapy Oncology Group. Int J Radiat Oncol Biol Phys. 2005;63(2):577–83. - PubMed
    1. Smith AR. Vision 20/20: proton therapy. Med Phys. 2009;36(2):556–68. - PubMed
    1. Schneider U, Pedroni E, Lomax A. The calibration of CT Hounsfield units for radiotherapy treatment planning. Phys Med Biol. 1996;41(1):111–24. - PubMed

Publication types

LinkOut - more resources