Background: The three softwares, Quantitative Perfusion SPECT (QPS), Emory Cardiac Toolbox, and 4 Dimension-Myocardial SPECT (4DM) are widely used with myocardial perfusion imaging (MPI) to determine perfusion defect size (PDS) and left ventricular (LV) function. There are limited data on the degree of agreement between these methods in quantifying the LV perfusion pattern and function.
Methods and results: In 120 consecutive patients who had abnormal regadenoson SPECT MPI with a visually derived summed stress score ≥4, the correlation between the softwares for measurements of PDS, reversible, and fixed defects was poor to fair (Spearman's ρ = 0.18-0.72). Overall, estimation of defect size was smaller by QPS and larger by 4DM. There was discordance among the softwares in 62% of the cases in defining PDS as small/moderate/large. The correlation between the softwares was better for measuring LVEF, volumes and mass (ρ = 0.84-0.97), and discrepant results for defining normal/mild-moderate/severe LV systolic dysfunction were prevalent in 28% of the patients.
Conclusion: There are significant differences between the softwares in measuring PDS as well as LV function, and more importantly in defining small, moderate, or large ischemic burden. These results suggest the necessity of using the same software when assessing interval changes by serial imaging.
Comment in
-
Call for a standard unit of defect size.J Nucl Cardiol. 2014 Aug;21(4):682-4. doi: 10.1007/s12350-014-9874-8. Epub 2014 Apr 9. J Nucl Cardiol. 2014. PMID: 24715620 No abstract available.
-
Thoughts on Wackers' call for a standard unit of defect size.J Nucl Cardiol. 2014 Aug;21(4):685. doi: 10.1007/s12350-014-9875-7. Epub 2014 Apr 9. J Nucl Cardiol. 2014. PMID: 24715621 No abstract available.
Similar articles
-
Assessment of four different cardiac softwares for evaluation of LVEF with CZT-SPECT vs CMR in 48 patients with recent STEMI.J Nucl Cardiol. 2020 Dec;27(6):2017-2026. doi: 10.1007/s12350-018-01493-y. Epub 2018 Nov 13. J Nucl Cardiol. 2020. PMID: 30426398
-
Evaluation of changes in perfusion defect and left ventricular systolic function using Tc-99m Tetrofosmin single photon emission computed tomography over 3 month period in patients of Acute Myocardial Infarction undergoing primary angioplasty.Nucl Med Rev Cent East Eur. 2018;21(1):1-7. doi: 10.5603/NMR.a2018.0001. Epub 2018 Jan 10. Nucl Med Rev Cent East Eur. 2018. PMID: 29319132
-
Evaluation of Left Ventricular Volumes and Ejection Fraction from Gated Myocardial Perfusion SPECT Processed with "Myovation Evolution": Comparison of Three Automated Software Packages using Cardiac Magnetic Resonance as Reference.Curr Radiopharm. 2021;14(2):112-120. doi: 10.2174/1874471013666200915130100. Curr Radiopharm. 2021. PMID: 32933466
-
Serial myocardial perfusion imaging: defining a significant change and targeting management decisions.JACC Cardiovasc Imaging. 2014 Jan;7(1):79-96. doi: 10.1016/j.jcmg.2013.05.022. JACC Cardiovasc Imaging. 2014. PMID: 24433711 Review.
-
Artifacts in Quantitative analysis of myocardial perfusion SPECT, using Cedars-Sinai QPS Software.J Nucl Cardiol. 2017 Apr;24(2):534-542. doi: 10.1007/s12350-016-0726-6. Epub 2016 Nov 10. J Nucl Cardiol. 2017. PMID: 27834038 Review.
Cited by
-
The effect of CT-based attenuation correction on the automatic perfusion score of myocardial perfusion imaging using a dedicated cardiac solid-state CZT SPECT/CT.J Nucl Cardiol. 2019 Feb;26(1):236-245. doi: 10.1007/s12350-017-0905-0. Epub 2017 May 1. J Nucl Cardiol. 2019. PMID: 28462467
-
Serial imaging and outcome prediction.J Nucl Cardiol. 2016 Feb;23(1):117-21. doi: 10.1007/s12350-015-0312-3. Epub 2015 Oct 30. J Nucl Cardiol. 2016. PMID: 26514576 Review. No abstract available.
-
Towards Standardization of Quantitative Retinal Vascular Parameters: Comparison of SIVA and VAMPIRE Measurements in the Lothian Birth Cohort 1936.Transl Vis Sci Technol. 2018 Mar 23;7(2):12. doi: 10.1167/tvst.7.2.12. eCollection 2018 Mar. Transl Vis Sci Technol. 2018. PMID: 29600120 Free PMC article.
-
Subacute cardiac rubidium-82 positron emission tomography (82Rb-PET) to assess myocardial area at risk, final infarct size, and myocardial salvage after STEMI.J Nucl Cardiol. 2018 Jun;25(3):970-981. doi: 10.1007/s12350-016-0694-x. Epub 2016 Oct 14. J Nucl Cardiol. 2018. PMID: 27743299 Free PMC article.
-
Review of cardiovascular imaging in the Journal of Nuclear Cardiology 2018. Part 1 of 2: Positron emission tomography, computed tomography, and magnetic resonance.J Nucl Cardiol. 2019 Apr;26(2):524-535. doi: 10.1007/s12350-018-01558-y. Epub 2019 Jan 2. J Nucl Cardiol. 2019. PMID: 30603892 Review.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous