222Rn+220Rn monitoring by alpha spectrometry
- PMID: 24723194
- DOI: 10.1093/rpd/ncu071
222Rn+220Rn monitoring by alpha spectrometry
Abstract
Controlled 222Rn+220Rn mixed atmospheres have been realised introducing calibrated sources in a stainless steel chamber. An electrostatic alpha monitor internal to the chamber has been used for an accurate discrimination of alpha peaks due to the products of the two isotopes. In the chamber, different specific activities are achieved in order to test the response of the internal reference instrument and to evaluate the possible interferences due to contemporary presence of both radon isotopes. Results show that: (i) the atmospheres are very stable, (ii) the monitor is adequate for their control because the various alpha lines are well evaluated and (iii) using Tyvek® filter, the efficiency of monitor is stable and constant vs. activity.
© The Author 2014. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.