High-quality image acquisition by double exposure overlap in dental cone beam computed tomography
- PMID: 24736110
- DOI: 10.1016/j.oooo.2014.02.024
High-quality image acquisition by double exposure overlap in dental cone beam computed tomography
Abstract
Objective: With a double exposure overlapping cone beam computed tomography (CBCT) scan technique, using CBCT acquisition radiation dose, the objective was to obtain apparent density similar to that of multidetector computed tomography (MDCT).
Study design: Factory quality-assurance phantom and water phantom were used for the evaluation of apparent density fidelity of iCAT scans in different modes. Each scan's apparent density was analyzed for identical regions using ImageJ, version 1.42q.
Results: The iCAT Classic extended height acquisition with 4-cm central overlap and reconstruction of 2 groups of 300 projections per rotation for the water and quality-assurance CBCT phantoms resulted in improved apparent density fidelity. This apparent density accuracy was superior to that of iCAT scan at high resolution (600 projections during 1 rotation).
Conclusions: Using double exposure overlapping CBCT scans allows the analysis quality to be comparable with that of MDCT.
Copyright © 2014 Elsevier Inc. All rights reserved.
Similar articles
-
Performance of cone-beam computed tomography and multidetector computed tomography in diagnostic imaging of the midface: A comparative study on Phantom and cadaver head scans.Eur Radiol. 2017 Feb;27(2):790-800. doi: 10.1007/s00330-016-4387-2. Epub 2016 May 11. Eur Radiol. 2017. PMID: 27169574
-
The effect of calibration and detector temperature on the reconstructed cone beam computed tomography image quality: a study for the workflow of the iCAT Classic equipment.Oral Surg Oral Med Oral Pathol Oral Radiol. 2015 Apr;119(4):473-80. doi: 10.1016/j.oooo.2014.12.009. Epub 2014 Dec 31. Oral Surg Oral Med Oral Pathol Oral Radiol. 2015. PMID: 25687195
-
[Analysis of image quality and dose calculation accuracy in cone beam CT acquisitions with limited projection data (half scan, half fan) with regard to usability for adaptive radiation therapy treatment planning].Z Med Phys. 2011;21(1):11-8. doi: 10.1016/j.zemedi.2010.08.002. Z Med Phys. 2011. PMID: 20888201 German.
-
[Cone Beam computerized tomography and implants].Rev Stomatol Chir Maxillofac. 2012 Sep;113(4):245-58. doi: 10.1016/j.stomax.2012.05.009. Epub 2012 Aug 11. Rev Stomatol Chir Maxillofac. 2012. PMID: 22889565 Review. French.
-
Are multi-detector computed tomography and cone-beam computed tomography exams and software accurate to measure the upper airway? A systematic review.Eur J Orthod. 2023 Nov 30;45(6):818-831. doi: 10.1093/ejo/cjad060. Eur J Orthod. 2023. PMID: 37797294
Cited by
-
CBCT-based bone quality assessment: are Hounsfield units applicable?Dentomaxillofac Radiol. 2015;44(1):20140238. doi: 10.1259/dmfr.20140238. Dentomaxillofac Radiol. 2015. PMID: 25315442 Free PMC article. Review.
-
Change in Image Quality According to the 3D Locations of a CBCT Phantom.PLoS One. 2016 Apr 19;11(4):e0153884. doi: 10.1371/journal.pone.0153884. eCollection 2016. PLoS One. 2016. PMID: 27093639 Free PMC article.
-
Validation of bone mineral density measurement using quantitative CBCT image based on deep learning.Sci Rep. 2023 Jul 24;13(1):11921. doi: 10.1038/s41598-023-38943-8. Sci Rep. 2023. PMID: 37488135 Free PMC article.
-
Comparison of the Hounsfield unit in CT scan with the gray level in cone-beam CT.J Dent Res Dent Clin Dent Prospects. 2019 Summer;13(3):177-182. doi: 10.15171/joddd.2019.028. Epub 2019 Oct 7. J Dent Res Dent Clin Dent Prospects. 2019. PMID: 31857863 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources