Comparison of continuous non-invasive finger arterial pressure monitoring with conventional intermittent automated arm arterial pressure measurement in patients under general anaesthesia
- PMID: 24740992
- DOI: 10.1093/bja/aeu091
Comparison of continuous non-invasive finger arterial pressure monitoring with conventional intermittent automated arm arterial pressure measurement in patients under general anaesthesia
Abstract
Background: For a majority of patients undergoing anaesthesia for general surgery, mean arterial pressure (MAP) is only measured intermittently by arm cuff oscillometry (MAPiNIAP). In contrast, the Nexfin(®) device provides continuous non-invasive measurement of MAP (MAPcNIAP) using a finger cuff. We explored the agreement of MAPcNIAP and MAPiNIAP with the gold standard: continuous invasive MAP measurement by placement of a radial artery catheter (MAPinvasive).
Methods: In a total of 120 patients undergoing elective general surgery and clinically requiring MAPinvasive measurement, MAPiNIAP and MAPcNIAP were measured in a 30 min time period at an arbitrary moment during surgery with stable haemodynamics. MAPiNIAP was measured every 5 min.
Results: Data from 112 patients were analysed. Compared with MAPinvasive, modified Bland-Altman analysis revealed a bias (sd) of 2 (9) mm Hg for MAPcNIAP and -2 (12) mm Hg for MAPiNIAP. Percentage errors for MAPcNIAP and MAPiNIAP were 22% and 32%, respectively.
Conclusions: In a haemodynamically stable phase in patients undergoing general anaesthesia, the agreement with invasive MAP of continuous non-invasive measurement using a finger cuff was not inferior to the agreement of intermittent arm cuff oscillometry. Continuous measurements using a finger cuff can interchangeably be used as an alternative for intermittent arm cuff oscillometry in haemodynamically stable patients, with the advantage of beat-to-beat haemodynamic monitoring.
Clinical trial registration: NCT 01362335 (clinicaltrials.gov).
Trial registration: ClinicalTrials.gov NCT01362335.
Keywords: NIAP; Nexfin; non-invasive arterial pressure; volume clamp.
© The Author [2014]. Published by Oxford University Press on behalf of the British Journal of Anaesthesia. All rights reserved. For Permissions, please email: journals.permissions@oup.com.
Comment in
-
Are we ready for the age of non-invasive haemodynamic monitoring?Br J Anaesth. 2014 Sep;113(3):340-3. doi: 10.1093/bja/aeu145. Epub 2014 May 31. Br J Anaesth. 2014. PMID: 24880826 No abstract available.
-
Are we ready for non-invasive blood pressure monitoring?Br J Anaesth. 2015 Jul;115(1):130. doi: 10.1093/bja/aev184. Br J Anaesth. 2015. PMID: 26089449 No abstract available.
-
Are we ready for non-invasive blood pressure monitoring?—reply.Br J Anaesth. 2015 Jul;115(1):130-1. doi: 10.1093/bja/aev185. Br J Anaesth. 2015. PMID: 26089450 No abstract available.
Similar articles
-
Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurement.Br J Anaesth. 2012 Feb;108(2):202-10. doi: 10.1093/bja/aer394. Epub 2011 Dec 12. Br J Anaesth. 2012. PMID: 22171358 Clinical Trial.
-
Precision and accuracy of a new device (CNAPTM) for continuous non-invasive arterial pressure monitoring: assessment during general anaesthesia.Br J Anaesth. 2010 Sep;105(3):264-72. doi: 10.1093/bja/aeq143. Epub 2010 Jul 13. Br J Anaesth. 2010. PMID: 20627878
-
Use of the Nexfin™ device to detect acute arterial pressure variations during anaesthesia induction.Br J Anaesth. 2014 Jul;113(1):52-60. doi: 10.1093/bja/aeu055. Epub 2014 Apr 25. Br J Anaesth. 2014. PMID: 24771806 Clinical Trial.
-
[Intra-operative continuous non-invasive blood pressure monitoring].Ugeskr Laeger. 2018 Sep 24;180(39):V01180007. Ugeskr Laeger. 2018. PMID: 30274570 Review. Danish.
-
The accuracy of noninvasive cardiac output and pressure measurements with finger cuff: a concise review.Curr Opin Crit Care. 2015 Jun;21(3):232-9. doi: 10.1097/MCC.0000000000000198. Curr Opin Crit Care. 2015. PMID: 25922896 Review.
Cited by
-
Haemodynamic monitoring during noncardiac surgery: past, present, and future.J Clin Monit Comput. 2024 Jun;38(3):565-580. doi: 10.1007/s10877-024-01161-2. Epub 2024 Apr 30. J Clin Monit Comput. 2024. PMID: 38687416 Free PMC article. Review.
-
Hemodynamics and tissue oxygenation during balanced anesthesia with a high antinociceptive contribution: an observational study.Perioper Med (Lond). 2014 Oct 30;3:9. doi: 10.1186/2047-0525-3-9. eCollection 2014. Perioper Med (Lond). 2014. PMID: 25632346 Free PMC article.
-
The future of intraoperative blood pressure management.J Clin Monit Comput. 2018 Feb;32(1):1-4. doi: 10.1007/s10877-017-9989-y. Epub 2017 Feb 7. J Clin Monit Comput. 2018. PMID: 28168583 No abstract available.
-
Comparison of consistency between wireless and conventional wired monitoring systems in gastrointestinal endoscopy.Sci Rep. 2025 Aug 3;15(1):28296. doi: 10.1038/s41598-025-12927-2. Sci Rep. 2025. PMID: 40754611 Free PMC article.
-
ClearSight™ finger cuff versus invasive arterial pressure measurement in patients with body mass index above 45 kg/m2.BMC Anesthesiol. 2021 May 18;21(1):152. doi: 10.1186/s12871-021-01374-x. BMC Anesthesiol. 2021. PMID: 34006231 Free PMC article.
Publication types
MeSH terms
Associated data
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical