. 2014 Jul;25(7):781-6.
doi: 10.1111/jce.12437. Epub 2014 Jun 3.

Compatibility of electroanatomical mapping systems with a concurrent percutaneous axial flow ventricular assist device

Affiliations

Compatibility of electroanatomical mapping systems with a concurrent percutaneous axial flow ventricular assist device

Vaibhav R Vaidya et al. J Cardiovasc Electrophysiol. 2014 Jul.

Abstract

Background: Hemodynamic instability hinders activation and entrainment mapping during ventricular tachycardia ablation. The Impella 2.5 microaxial flow device (MFD; Abiomed Inc., Danvers, MA, USA) is used to prevent hemodynamic instability during electrophysiologic study. However, electromagnetic interference (EMI) generated by this device can preclude accurate electroanatomic mapping.

Methods: Impella was placed in the left ventricle of 7 canines for circulatory support. Electroanatomic mapping during sinus rhythm, ventricular pacing, and ventricular fibrillation (VF) was performed using magnet- (CARTO3, Biosense Webster Inc., Diamond Bar, CA, USA) and impedance- (EnSite Velocity System/EnSite NavX, St. Jude Medical Inc., St. Paul, MN, USA) based systems. Distance from device to points with severe EMI precluding acquisition was compared to points with mild/no EMI. Two methods were used to reduce EMI: (1) titration of MFD performance, and (2) impedance-only mapping combined with manual annotation of activation.

Results: Severe EMI did not occur during impedance-based mapping. Severe EMI was observed using CARTO3 at 9.4% of all points attempted at maximum performance level (P8) of device. Severe EMI occurred at points closer to device (40.1 ± 16.8 mm) versus (55.5 ± 20.0 mm) for mild/no EMI, P < 0.0001. Severe EMI using CARTO3 was resolved by either (1) reduction of performance from P8 to P6 or (2) impedance-only mapping with manual annotation.

Conclusion: Concurrent use of MFD caused EMI to prevent acquisition of points with magnet-based mapping. Predictors for EMI were distance from device and performance level. Temporary reductions to P6 or impedance-only mapping are 2 methods to resolve EMI.

Keywords: Impella; catheter ablation; electroanatomical mapping systems; impedance; magnetic; ventricular tachycardia.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Right (a) and left (b) ventricular endocardial mapping, epicardial mapping (c) in sinus rhythm with the MFD at maximal performance using the magnet-based mapping system. Blue points: severe distortion, orange points: mild distortion, green points: no distortion. Endocardial maps (a, b) have greater numbers of blue dots compared with the epicardial map (c); indicating greater occurrence of severe distortion with endocardial maps.

References

    1. Bansch D, Oyang F, Antz M, Arentz T, Weber R, Val-Mejias JE, Ernst S, Kuck KH. Successful catheter ablation of electrical storm after myocardial infarction. Circulation. 2003;108:3011–3016. - PubMed
    1. Moss AJ, Greenberg H, Case RB, Zareba W, Hall WJ, Brown MW, Daubert JP, McNitt S, Andrews ML, Elkin AD. Long-term clinical course of patients after termination of ventricular tachyarrhythmia by an implanted defibrillator. Circulation. 2004;110:3760–3765. - PubMed
    1. Reddy VY, Reynolds MR, Neuzil P, Richardson AW, Taborsky M, Jongnarangsin K, Kralovec S, Sediva L, Ruskin JN, Josephson ME. Prophylactic catheter ablation for the prevention of defibrillator therapy. N Engl J Med. 2007;357:2657–2665. - PMC - PubMed
    1. Stevenson WG, Soejima K. Catheter ablation for ventricular tachycardia. Circulation. 2007;115:2750–2760. - PubMed
    1. Yamada T, Kay GN. Optimal ablation strategies for different types of ventricular tachycardias. Nat Rev Cardiol. 2012;9:512–525. - PubMed