Is portable three-dimensional ultrasound a valid technique for measurement of postpartum urinary bladder volume?
- PMID: 24767639
- DOI: 10.1016/j.tjog.2013.01.028
Is portable three-dimensional ultrasound a valid technique for measurement of postpartum urinary bladder volume?
Abstract
Objective: To determine the accuracy of bladder volume (BV) measurement and to identify factors that influenced the ultrasound bladder scanner (UBS) measurement of BV in postpartum women compared with transurethral catheterization.
Materials, methods and results: A total of 190 paired measurements were performed on 190 women aged between 16 and 47 years. Majority of the women (36.9%) were in their first parity (range: 1-9). The mean BV was 159.46 mL [standard deviation (SD) = 99.78; range: 17-593 mL] on three-dimensional (3D) UBS and 143.76 mL (SD = 104.89; range: 2-588 mL) on catheterization. The mean difference between the two values was 15.70 mL (SD = 69.31, p < 0.001). The Foley's catheter was clamped for a minimum of 30 minutes and a maximum of 260 minutes. The UBS measurements were highly correlated with the BV obtained by catheterization (r = 0.819; p < 0.001). There was no significant difference between UBS and catheterized volume in a different volume subgroup. There were 35 cases (18.2%) with the difference in measurement of more than 25%. Twenty-five cases were associated with a catheterized volume less than 50 mL. There were a total of 33 cases with true BV of less than 50 mL, and only eight cases (24.24%) had an accurate measurement. The UBS tends to overestimate BV between 2 and 270 mL in this group. There was a significant linear relationship between the difference in measurement and body weight and duration of clamping.
Conclusion: Measurement of BV in immediate postpartum period using UBS is comparable with urethral catheterization. The accuracy of measurement is affected by body weight and increasing amount of BV.
Keywords: postpartum; three-dimensional ultrasound; urinary bladder volume.
Copyright © 2014. Published by Elsevier B.V.
Similar articles
-
Accuracy of Post-Void Residual Urine Volume Measurement Using an Ultrasound Bladder Scanner among Postoperative Radical Hysterectomy Patients.J Med Assoc Thai. 2016 Oct;99(10):1061-6. J Med Assoc Thai. 2016. PMID: 29952185
-
Accuracy and precision of a new portable ultrasound scanner, the BME-150A, in residual urine volume measurement: a comparison with the BladderScan BVI 3000.Int Urogynecol J Pelvic Floor Dysfunct. 2007 Jun;18(6):641-4. doi: 10.1007/s00192-006-0219-2. Epub 2006 Nov 7. Int Urogynecol J Pelvic Floor Dysfunct. 2007. PMID: 17115233 Clinical Trial.
-
Determining the reliability of ultrasound measurements and the validity of the formulae for ultrasound estimation of postvoid residual bladder volume in postpartum women.Neurourol Urodyn. 2003;22(3):255-60. doi: 10.1002/nau.10112. Neurourol Urodyn. 2003. PMID: 12707878
-
The effectiveness of the ultrasound bladder scanner in reducing urinary tract infections: a meta-analysis.J Clin Nurs. 2010 Nov;19(21-22):2970-9. doi: 10.1111/j.1365-2702.2010.03281.x. Epub 2010 Sep 8. J Clin Nurs. 2010. PMID: 21040003 Review.
-
[Efficacy and problems of bladder volume measurement using portable three dimensional ultrasound scanning device--in particular, on measuring bladder volume lower than 100ml].Nihon Hinyokika Gakkai Zasshi. 2005 Sep;96(6):601-9. doi: 10.5980/jpnjurol1989.96.601. Nihon Hinyokika Gakkai Zasshi. 2005. PMID: 16218402 Review. Japanese.
Cited by
-
Feasibility of a novel 3-dimensional mathematical algorithmic computation of feline bladder volumes using point-of-care longitudinal and transverse cysto-colic ultrasonographic images.Can J Vet Res. 2019 Oct;83(4):298-312. Can J Vet Res. 2019. PMID: 31571731 Free PMC article.
-
Postpartum urinary retention: what are the sequelae? A long-term study and review of the literature.Int Urogynecol J. 2022 Jun;33(6):1601-1608. doi: 10.1007/s00192-021-05074-5. Epub 2022 Feb 7. Int Urogynecol J. 2022. PMID: 35129645 Free PMC article. Review.
-
Urinary Retention in Pregnancy and Puerperium: Acupuncture Treatment.Med Acupunct. 2019 Oct 1;31(5):269-273. doi: 10.1089/acu.2019.1363. Epub 2019 Oct 17. Med Acupunct. 2019. PMID: 31624525 Free PMC article. Review.
-
Comparison of 2D and 3D ultrasound methods to measure serial bladder volumes during filling: Steps toward development of non-invasive ultrasound urodynamics.Bladder (San Franc). 2018;5(1):e32. doi: 10.14440/bladder.2018.565. Epub 2018 Jan 4. Bladder (San Franc). 2018. PMID: 29354653 Free PMC article.
-
Accuracy of postvoid residual volumes after vaginal delivery: a prospective equivalence study to compare an automatic scanning device with transurethral catheterization.Int Urogynecol J. 2019 May;30(5):773-778. doi: 10.1007/s00192-018-3700-9. Epub 2018 Jun 27. Int Urogynecol J. 2019. PMID: 29951911 Free PMC article. Clinical Trial.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical