Review
. 2014 Mar;5(Suppl 1):S17-20.

Strategies to reduce pitfalls in measuring blood pressure

Affiliations
Review

Strategies to reduce pitfalls in measuring blood pressure

Hamidreza Badeli et al. Int J Prev Med. 2014 Mar.

Abstract

Errors in blood pressure (BP) measurement are common in the clinical practice. Inaccurate measurements of BP may lead to misdiagnosis and inappropriate treatment of hypertension. The preferred method of BP measurement in the clinical setting is auscultation, using the first and the fifth Korotkoff sounds. However, the use of mercury sphygmomanometer is declining. Automated oscillometric devices are an acceptable alternative method of BP measurements if the proper cuff size is used. Aneroid devices are suitable, but they require frequent calibration. There is increasing evidence that home readings predict cardiovascular events and are particularly useful for monitoring the effects of treatment. At 24 h ambulatory monitoring is also useful for diagnosing white-coat hypertension and resistance hypertension. There is increasing evidence that lack of nocturnal BP dipping during the night may be associated with increased cardiovascular event. This report attempts to address the need for accurate BP measurements in children and adolescents by reducing human and equipment errors and providing clinicians with the accurate measurement of BP, which is essential to classify individuals, to ascertain BP-related CV risks and to guide management.

Keywords: Blood pressure measurement; children; method; pitfalls.

PubMed Disclaimer

Conflict of interest statement

Conflict of Interest: None declared

Figures

Figure 1
Figure 1
Technique of blood pressure measurement

Similar articles

Cited by

References

    1. Flynn JT, Falkner BE. The importance of blood pressure screening in children. J Pediatr. 2009;155:299. - PubMed
    1. Podoll A, Grenier M, Croix B, Feig DI. Inaccuracy in pediatric outpatient blood pressure measurement. Pediatrics. 2007;119:e538–43. - PubMed
    1. Prineas RJ, Ostchega Y, Carroll M, Dillon C, McDowell M. US demographic trends in mid-arm circumference and recommended blood pressure cuffs for children and adolescents: Data from the National Health and Nutrition Examination Survey 1988-2004. Blood Press Monit. 2007;12:75–80. - PubMed
    1. Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN, et al. Recommendations for blood pressure measurement in humans and experimental animals: Part 1: Blood pressure measurement in humans: A statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Hypertension. 2005;45:142–61. - PubMed
    1. Mitchell PL, Parlin RW, Blackburn H. effect of vertical displacement of the arm on indirect blood-pressure measurement. N Engl J Med. 1964;271:72–4. - PubMed