Comparative Study
. 2014 May 2;4(5):e004565.
doi: 10.1136/bmjopen-2013-004565.

Triage tests for identifying atrial fibrillation in primary care: a diagnostic accuracy study comparing single-lead ECG and modified BP monitors

Affiliations
Comparative Study

Triage tests for identifying atrial fibrillation in primary care: a diagnostic accuracy study comparing single-lead ECG and modified BP monitors

Karen Kearley et al. BMJ Open. .

Abstract

Objective: New electronic devices offer an opportunity within routine primary care settings for improving the detection of atrial fibrillation (AF), which is a common cardiac arrhythmia and a modifiable risk factor for stroke. We aimed to assess the performance of a modified blood pressure (BP) monitor and two single-lead ECG devices, as diagnostic triage tests for the detection of AF.

Setting: 6 General Practices in the UK.

Participants: 1000 ambulatory patients aged 75 years and over.

Primary and secondary outcome measures: Comparative diagnostic accuracy of modified BP monitor and single-lead ECG devices, compared to reference standard of 12-lead ECG, independently interpreted by cardiologists.

Results: A total of 79 participants (7.9%) had AF diagnosed by 12-lead ECG. All three devices had a high sensitivity (93.9-98.7%) and are useful for ruling out AF. WatchBP is a better triage test than Omron autoanalysis because it is more specific-89.7% (95% CI 87.5% to 91.6%) compared to 78.3% (95% CI 73.0% to 82.9%), respectively. This would translate into a lower follow-on ECG rate of 17% to rule in/rule out AF compared to 29.7% with the Omron text message in the study population. The overall specificity of single-lead ECGs analysed by a cardiologist was 94.6% for Omron and 90.1% for Merlin.

Conclusions: WatchBP performs better as a triage test for identifying AF in primary care than the single-lead ECG monitors as it does not require expertise for interpretation and its diagnostic performance is comparable to single-lead ECG analysis by cardiologists. It could be used opportunistically to screen elderly patients for undiagnosed AF at regular intervals and/or during BP measurement.

Keywords: Primary Care; Stroke Medicine.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Patient recruitment flowchart.
Figure 2
Figure 2
Predictive value of WatchBP and Omron autoanalysis for detecting atrial fibrillation (AF).
Figure 3
Figure 3
Forest plot of diagnostic accuracy per cardiologist of Omron and Merlin single-lead ECGs.

References

    1. NICE. Atrial fibrillation: national clinical guideline for management in primary and secondary care. Royal College of Physicians, 2006 - PubMed
    1. Fitzmaurice DA, Hobbs FDR, Jowett S, et al. Screening versus routine practice for detection of atrial fibrillation in people aged 65 or over: cluster randomised controlled trial. BMJ 2007;335: 383–6 - PMC - PubMed
    1. Miyasaka Y, Barnes ME, Bailey KR, et al. Mortality trends in patients diagnosed with first atrial fibrillation—a 21-year community study. J Am Coll Cardiol 2007;49:986–92 - PubMed
    1. Wolf PA, Abbott RD, Kannell WB. Atrial fibrillation as an independent risk factor for stroke: the Framingham study. Stroke 1991;22: 983–8 - PubMed
    1. Aguilar MI, Hart R. Oral anticoagulants for preventing stroke in patients with non-valvular atrial fibrillation and no previous history of stroke or transient ischemic attacks. Cochrane Database Syst Rev 2005;(3):CD001927. - PMC - PubMed

Publication types

LinkOut - more resources