Dose reduction for coronary calcium scoring with hybrid and model-based iterative reconstruction: an ex vivo study
- PMID: 24794292
- DOI: 10.1007/s10554-014-0434-8
Dose reduction for coronary calcium scoring with hybrid and model-based iterative reconstruction: an ex vivo study
Abstract
Purpose: To determine the influence of dose reduction on coronary calcium scoring using hybrid and model-based iterative reconstruction (IR) techniques.
Methods: Fifteen ex vivo hearts were scanned in a phantom representing an average adult person at routine dose and three levels of dose reduction; 27, 55 and 82% reduced-dose, respectively. All images were reconstructed using filtered back-projection (FBP), hybrid IR (iDose4, levels 1, 4 and 7) as well as model-based IR iterative model reconstruction (IMR, levels 1, 2 and 3). Agatston, mass and volume scores found with iDose4 and IMR were compared to FBP reconstruction (routine dose) as well as objective image quality.
Results: With FBP calcium scores remained unchanged at 82% reduced dose. With IR Agatston scores differed significantly at routine dose, using IMR level 3 and iDose4 level 7, and at 82% reduced dose, using IMR levels 1-3 and iDose4 level 7. The maximum median difference was 5.3%. Mass remained unchanged at reduced dose levels while volume was significantly lower at 82% reduced dose with IMR (maximum median difference 5.0%). Objective image quality improved with IR, at 82% reduced dose the CNR of iDose4 level 7 was similar to the reference dose CNR, and IMR levels 1-3 resulted in an even higher CNR.
Conclusion: Calcium scores were not affected by radiation-dose reduction with FBP and low levels of hybrid IR. Objective image quality increased significantly using hybrid and model-based IR. Therefore low level hybrid IR has the potential to reduce radiation-dose of coronary calcium scoring with up to 82%.
Similar articles
-
The effect of iterative model reconstruction on coronary artery calcium quantification.Int J Cardiovasc Imaging. 2016 Jan;32(1):153-60. doi: 10.1007/s10554-015-0740-9. Epub 2015 Aug 19. Int J Cardiovasc Imaging. 2016. PMID: 26285899
-
Value of knowledge-based iterative model reconstruction in low-kV 256-slice coronary CT angiography.J Cardiovasc Comput Tomogr. 2014 Mar-Apr;8(2):115-23. doi: 10.1016/j.jcct.2013.12.010. Epub 2014 Jan 12. J Cardiovasc Comput Tomogr. 2014. PMID: 24661824
-
Prospective evaluation of the influence of iterative reconstruction on the reproducibility of coronary calcium quantification in reduced radiation dose 320 detector row CT.J Cardiovasc Comput Tomogr. 2016 Sep-Oct;10(5):359-63. doi: 10.1016/j.jcct.2016.07.016. Epub 2016 Jul 27. J Cardiovasc Comput Tomogr. 2016. PMID: 27591767 Free PMC article. Clinical Trial.
-
Radiation dose reduction for coronary artery calcium scoring at 320-detector CT with adaptive iterative dose reduction 3D.Int J Cardiovasc Imaging. 2015 Jun;31(5):1045-52. doi: 10.1007/s10554-015-0637-7. Epub 2015 Mar 10. Int J Cardiovasc Imaging. 2015. PMID: 25754302
-
Iterative beam-hardening correction with advanced modeled iterative reconstruction in low voltage CT coronary calcium scoring with tin filtration: Impact on coronary artery calcium quantification and image quality.J Cardiovasc Comput Tomogr. 2017 Sep-Oct;11(5):354-359. doi: 10.1016/j.jcct.2017.07.003. Epub 2017 Jul 24. J Cardiovasc Comput Tomogr. 2017. PMID: 28756086
Cited by
-
Accuracy of coronary artery calcium scoring with tube current reduction by 75%, using an adaptive iterative reconstruction algorithm.Br J Radiol. 2018 Apr;91(1084):20170678. doi: 10.1259/bjr.20170678. Epub 2018 Jan 10. Br J Radiol. 2018. PMID: 29260893 Free PMC article.
-
Computed tomography image quality of aortic stents in patients with aortic coarctation: a multicentre evaluation.Eur Radiol Exp. 2018 Jul 18;2(1):17. doi: 10.1186/s41747-018-0046-5. Eur Radiol Exp. 2018. PMID: 33252748 Free PMC article.
-
In vitro evaluation of a new iterative reconstruction algorithm for dose reduction in coronary artery calcium scoring.Acta Radiol Open. 2017 May 29;6(5):2058460117710682. doi: 10.1177/2058460117710682. eCollection 2017 May. Acta Radiol Open. 2017. PMID: 28607763 Free PMC article.
-
Influence of dose reduction and iterative reconstruction on CT calcium scores: a multi-manufacturer dynamic phantom study.Int J Cardiovasc Imaging. 2017 Jun;33(6):899-914. doi: 10.1007/s10554-017-1061-y. Epub 2017 Jan 19. Int J Cardiovasc Imaging. 2017. PMID: 28102510 Free PMC article.
-
Determination of optimal imaging settings for urolithiasis CT using filtered back projection (FBP), statistical iterative reconstruction (IR) and knowledge-based iterative model reconstruction (IMR): a physical human phantom study.Br J Radiol. 2016;89(1058):20150527. doi: 10.1259/bjr.20150527. Epub 2015 Nov 18. Br J Radiol. 2016. PMID: 26577542 Free PMC article.