Technology and new fluorescence flow cytometry parameters in hematological analyzers
- PMID: 24797912
- PMCID: PMC6807107
- DOI: 10.1002/jcla.21747
Technology and new fluorescence flow cytometry parameters in hematological analyzers
Abstract
Background: Modern hematological analyzers offer a series of new tests in addition to the differential blood cell count. Here, we provide a short overview of the technique and the new tests included in the hematological analyzers manufactured by the company Sysmex (Kobe, Japan). As introduced here, the resulting new extended blood cell count allows for the quantification of DNA- and RNA-rich, activated and/or precursor leukocyte subpopulations.
Methods: To compare the different methods of blood cell counting, repetitive cell counting is carried out by hematological experts using microscopy and using the XE-5000 hematological analyzer.
Results: The hematological analyzer delivers more-accurate results with respect to normal cells but microscopy is suited better to recognize and count pathological cells.
Conclusion: With the new hematological analyzers, several new hematological tests are made available as screening tools that were not possible previously using standard cell-counting procedures.
Keywords: hematology.
© 2014 Wiley Periodicals, Inc.
References
-
- Ansari‐Lari MA, Kickler TS, Borowitz M. Immature granulocyte measurement using the Sysmex XE‐2100. Relationship to infection and sepsis. Am J Clin Pathol 2003;120:795–799. - PubMed
-
- Furundarena JR, Araiz M, Uranga M, et al. The utility of the Sysmex XE‐2100 analyzer's NEUT‐X and NEUT‐Y parameters for detecting neutrophil dysplasia in myelodysplastic syndromes. Int J Lab Hematol 2010;32:360–366. - PubMed
-
- Fernandes B, Hamaguchi Y. Automated enumeration of immature granulocytes. Am J Clin Pathol 2007;128(3):454–463. - PubMed
-
- di Mario A, Garzia M, Leone F, Arcangeli A, Pagano L, Zini G. Immature platelet fraction (IPF) in hospitalized patients with neutrophilia and suspected bacterial infection. J Infect 2009;59(3):201–206. - PubMed
-
- Information of the Company Sysmex . 2009. Floureszenz Durchflußzytometrie in der Hämatologie (XE5000 Case Manager). Sysmex Austria, Xtra; p1–11.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical