Effect of the anatomic reservoir on low-flow oxygen delivery via nasal cannula: constant flow versus pulse flow with portable oxygen concentrator
- PMID: 24803496
- DOI: 10.4187/respcare.02878
Effect of the anatomic reservoir on low-flow oxygen delivery via nasal cannula: constant flow versus pulse flow with portable oxygen concentrator
Abstract
Background: The F(IO(2)) for a nasal cannula with constant flow (CF) depends on the anatomic reservoir (AR), which is affected by changes in frequency and end-expiratory flow. Conversely, pulse flow (PF) devices do not require the AR. The purpose of this study was to compare the F(IO(2)) delivered by a nasal cannula supplied by CF via oxygen tank with that delivered by PF delivered via portable oxygen concentrator. Hypotheses were (1) a lung model of COPD with non-zero end-expiratory flow decreases F(IO(2)) for CF more than for PF, and (2) CF and PF perform differently in terms of F(IO(2)) delivery, despite having equivalent settings.
Methods: Normal and COPD lung models were simulated (IngMar Medical ASL 5000) using published human data: normal: breathing frequency = 15 breaths/min, R(in) = 4 cm H2O · s · L(-1), R(out) = 4 cm H2O · s · L(-1), C = 60 mL · cm H2O(-1), tidal volume (VT) = 685 mL, P(max) = 11.95 cm H2O, increase = 33%, and release = 28; COPD: breathing frequency = 20 breaths/min, R(in) = 12 cm H2O · s · L(-1), R(out) = 25 cm H2O · s · L(-1), C = 66 mL · cm H2O(-1), VT = 685 mL, Pmax = 24.52 cm H2O, increase = 35%, and release = 23%. CF was 1-5 L/min. Portable oxygen concentrators used were Solo2 (Invacare), XPO2 (Invacare), FreeStyle (AirSep), Focus (AirSep), One G3 (Inogen), and LifeChoice ActivOx (Inova Labs).
Results: CF produced significantly higher F(IO(2)) at all settings for normal lungs but lower for COPD lungs compared with Solo2. COPD reduced the F(IO(2)) for CF but had a smaller variable effect for PF. Data show there is no equivalency between PF setting and CF rates for the portable oxygen concentrators tested.
Conclusions: CF oxygen delivery via a nasal cannula is significantly reduced by elimination of the AR in a model of COPD, yielding clinically important decreases in F(IO(2)). PF (delivered with a portable oxygen concentrator) is relatively unaffected. This study supports the recommendation that clinicians and caretakers should titrate the PF setting to each patient's unique oxygen requirements.
Similar articles
-
Performance comparison of 4 portable oxygen concentrators.Respir Care. 2010 Apr;55(4):433-42. Respir Care. 2010. PMID: 20406511
-
Comparison of pulsed versus continuous oxygen delivery using realistic adult nasal airway replicas.Int J Chron Obstruct Pulmon Dis. 2017 Aug 24;12:2559-2571. doi: 10.2147/COPD.S141976. eCollection 2017. Int J Chron Obstruct Pulmon Dis. 2017. PMID: 28883723 Free PMC article.
-
In Vitro-In Silico Comparison of Pulsed Oxygen Delivery From Portable Oxygen Concentrators Versus Continuous Flow Oxygen Delivery.Respir Care. 2019 Feb;64(2):117-129. doi: 10.4187/respcare.06359. Epub 2019 Jan 29. Respir Care. 2019. PMID: 30696754
-
High-flow oxygen administration by nasal cannula for adult and perinatal patients.Respir Care. 2013 Jan;58(1):98-122. doi: 10.4187/respcare.01941. Respir Care. 2013. PMID: 23271822 Review.
-
Use of Nasal High Flow in Stable COPD: Rationale and Physiology.COPD. 2017 Jun;14(3):346-350. doi: 10.1080/15412555.2017.1315715. Epub 2017 May 1. COPD. 2017. PMID: 28459282 Review.
Cited by
-
Impact of Oxygen Concentration Delivered via Nasal Cannula on Different Lung Conditions: A Bench Study.Healthcare (Basel). 2021 Sep 19;9(9):1235. doi: 10.3390/healthcare9091235. Healthcare (Basel). 2021. PMID: 34575009 Free PMC article.
-
Contemporary portable oxygen concentrators and diverse breathing behaviours -- a bench comparison.BMC Pulm Med. 2019 Nov 19;19(1):217. doi: 10.1186/s12890-019-0980-x. BMC Pulm Med. 2019. PMID: 31744499 Free PMC article.
-
Variability in low-flow oxygen delivery by nasal cannula evaluated in neonatal and infant airway replicas.Respir Res. 2022 Dec 8;23(1):333. doi: 10.1186/s12931-022-02260-x. Respir Res. 2022. PMID: 36482438 Free PMC article.
-
A bench evaluation of fraction of oxygen in air delivery and tidal volume accuracy in home care ventilators available for hospital use.J Thorac Dis. 2016 Dec;8(12):3639-3647. doi: 10.21037/jtd.2016.12.64. J Thorac Dis. 2016. PMID: 28149559 Free PMC article.
-
The application of a surgical face mask over different oxygen delivery devices; a crossover study of measured end-tidal oxygen concentrations.BMC Anesthesiol. 2022 Mar 7;22(1):62. doi: 10.1186/s12871-022-01602-y. BMC Anesthesiol. 2022. PMID: 35255824 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials