Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2014 Oct;37(10):1265-73.
doi: 10.1111/pace.12421. Epub 2014 May 9.

Implantable cardioverter defibrillator system with floating atrial sensing dipole: a single-center experience

Affiliations

Implantable cardioverter defibrillator system with floating atrial sensing dipole: a single-center experience

Matteo Iori et al. Pacing Clin Electrophysiol. 2014 Oct.

Abstract

Background: The concept of a single-lead dual-chamber implantable cardioverter defibrillator (ICD) with floating sensing atrial dipole has been proven safe and functional. We report a single-center experience with this ICD system; the major focus of the work is on the recorded atrial activation and its stability on a medium term follow-up.

Methods: Thirteen patients received a DX ICD (BIOTRONIK SE & Co, Berlin, Germany) with the Linox Smart S DX(ProMRI) ICD lead; the implantation data were reported. Daily P- and R-wave sensing amplitude was collected and followed up during 200 days; their coefficient of variance (CV) was calculated. In addition, all the atrial and ventricular high-rate episodes were analyzed.

Results: The total x-ray exposure time was 3.9 ± 1.8 minutes. The overall mean sensing was 4.2 ± 1.9 mV for P wave and 12.9 ± 4.5 mV for R wave. The CV was significantly higher for the P-wave amplitude than for the R-wave one (0.25 ± 0.11 vs 0.08 ± 0.06; P < 0.001). A total of 27 high ventricular rate episodes were recorded and correctly discriminated by the device. Fifty-six high atrial rate episodes were recorded, 49 were true arrhythmic events.

Conclusions: The single-lead ICD system with floating atrial dipole provides reliable atrial sensing amplitude over time. The physician, without the implantation of an additional lead, has the atrial information that may be used for the discrimination of supraventricular tachyarrhythmia/ventricular tachycardia, for the early detection of atrial fibrillation episodes and for the evaluation of changes in the patient's heart status.

Keywords: P wave; atrial fibrillation; atrial sensing; implantable cardioverter defibrillator; tachyarrhythmias.

PubMed Disclaimer

LinkOut - more resources