Positional convulsant syncope in a pacemaker patient following insulation break of the right ventricular lead

Affiliations
Case Reports

Positional convulsant syncope in a pacemaker patient following insulation break of the right ventricular lead

Mehdi Ben Lassoued et al. BMJ Case Rep. .

Abstract

In spite of the advances made in the technology of pacemakers which resulted in a decrease in the incidence of pacemaker lead fracture, the latter remains a potential complication of implanted pacemakers manufactured in the early days. In this report, we present a case of fracture of the unipolar electrode diagnosed by an emergency physician in a patient on a pacemaker for 10 years who presented to the emergency department with positional convulsant syncopes.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Critical monitoring: severe bradycardia with non-conducted atrial spikes followed by asystole.
Figure 2
Figure 2
Critical monitoring: spontaneous resumption of a ventricular escape rhythm (wide QRS) at 60 bpm with atrial spikes.
Figure 3
Figure 3
X-ray film performed at the patient's bed: a fractured ventricular lead at the level of the right thoracic outlet (left-handed patient).
Figure 4
Figure 4
Chest X-ray after implantation of a new pacemaker on the left side.

References

    1. Saha A, Tan J, Prendergast B. Pacemaker lead fracture. Heart 2003;89:783. - PMC - PubMed
    1. Alt E, Volker R, Blomer H. Lead fracture in pacemaker patients. Thorac Cardiovasc Surg 1987;35:101–4 - PubMed
    1. Magney JE, Flynn DM, Parsons JA, et al. Anatomical mechanisms explaining damage to pacemaker leads, defibrillator leads, and failure of central venous catheters adjacent to the sternoclavicular joint. Pacing Clin Electrophysiol 1993;16:445–57 - PubMed
    1. Deering JA, Pederson DN. Pacemaker lead fracture associated with weightlifting: a report of two cases. Mil Med 1993;158:833–4 - PubMed
    1. Shu-Hsuan C, Che-Kim T, Shih-Huang L. Fracture of a pacemaker lead. CMAJ 2009;181:769 - PMC - PubMed