Large variation in blood flow between left ventricular segments, as detected by adenosine stress dynamic CT perfusion
- PMID: 24842265
- DOI: 10.1111/cpf.12163
Large variation in blood flow between left ventricular segments, as detected by adenosine stress dynamic CT perfusion
Abstract
Background: Dynamic cardiac CT perfusion (CTP) is based on repeated imaging during the first-pass contrast agent inflow. It is a relatively new method that still needs validation.
Purpose: To evaluate the variation in adenosine stress dynamic CTP blood flow as compared to (99m) Tc SPECT. Secondarily, to compare manual and automatic segmentation.
Methods: Seventeen patients with manifest coronary artery disease were included. Nine were excluded from evaluation for various reasons. All patients were examined with dynamic stress CTP and stress/rest SPECT. CTP blood flow was compared with SPECT on a per segment basis. Results for manual and automated AHA segmentation were compared.
Results: CTP showed a positive correlation with SPECT, with correlation coefficients of 0·38 and 0·41 for manual and automatic segmentation, respectively (P<0·0001). There was no significant difference between the correlation coefficients of the manual and automated segmentation procedures (P = 0·75). The average per individual global CTP blood flow value for normal segments varied by a factor of 1·9 (manual and automatic segmentation). For the whole patient group, the CTP blood flow value in normal segments varied by a factor of 2·9/2·7 (manual/automatic segmentation). Within each patient, the average per segment blood flow in normal segments varied by a factor of 1·3-2·0/1·2-2·1 (manual/automatic segmentation).
Conclusion: A positive but rather weak correlation was found between CTP and (99m) Tc SPECT. Large variations in CTP blood flow suggest that a cut-off value for stress myocardial blood flow is inadequate to detect ischaemic segments. Dynamic CTP is hampered by a limited coverage.
Keywords: AHA segmentation; contrast agent; correlation; ischaemia; tetrofosmin.
© 2014 Scandinavian Society of Clinical Physiology and Nuclear Medicine. Published by John Wiley & Sons Ltd.