. 2014 Jun;61(6):1809-21.
doi: 10.1109/TBME.2014.2307698.

Design and evaluation of a quasi-passive knee exoskeleton for investigation of motor adaptation in lower extremity joints

Design and evaluation of a quasi-passive knee exoskeleton for investigation of motor adaptation in lower extremity joints

Kamran Shamaei et al. IEEE Trans Biomed Eng. 2014 Jun.

Abstract

In this study, we describe the mechanical design and control scheme of a quasi-passive knee exoskeleton intended to investigate the biomechanical behavior of the knee joint during interaction with externally applied impedances. As the human knee behaves much like a linear spring during the stance phase of normal walking gait, the exoskeleton implements a spring across the knee in the weight acceptance (WA) phase of the gait while allowing free motion throughout the rest of the gait cycle, accomplished via an electromechanical clutch. The stiffness of the device is able to be varied by swapping springs, and the timing of engagement/disengagement changed to accommodate different loading profiles. After describing the design and control, we validate the mechanical performance and reliability of the exoskeleton through cyclic testing on a mechanical knee simulator. We then describe a preliminary experiment on three healthy adults to evaluate the functionality of the device on both left and right legs. The kinetic and kinematic analyses of these subjects show that the exoskeleton assistance can partially/fully replace the function of the knee joint and obtain nearly invariant moment and angle profiles for the hip and ankle joints, and the overall knee joint and exoskeleton complex under the applied moments of the exoskeleton versus the control condition, implying that the subjects undergo a considerable amount of motor adaptation in their lower extremities to the exoskeletal impedances, and encouraging more in-depth future experiments with the device.

PubMed Disclaimer

Publication types

LinkOut - more resources