Spontaneous fracture of diaphyseal stem of S-ROM femoral prosthesis
- PMID: 24849635
- PMCID: PMC4039855
- DOI: 10.1136/bcr-2013-202813
Spontaneous fracture of diaphyseal stem of S-ROM femoral prosthesis
Abstract
We present two cases of spontaneous fractures of the S-ROM femoral stem prosthesis implanted by different surgeons within 5 years of implantation. Both the stems fractured in the mid-distal stem at the junction of the main body and the slotted portion. Both fractures affected the posterior tine only. Our aim in publication is to ensure that this is an isolated problem and not an under-reported phenomenon. We are not aware of any previous reports of spontaneous fracture of the distal stem.
2014 BMJ Publishing Group Ltd.
Figures
Similar articles
-
Fracture of a spline of the femoral stem in an ingrown modular hip implant.Hip Int. 2016 Jul 25;26(4):e35-8. doi: 10.5301/hipint.5000399. Hip Int. 2016. PMID: 27443223
-
[Clinical results in early and mid term of using the S-ROM femoral stem in revision].Zhongguo Gu Shang. 2017 Apr 25;30(4):322-328. doi: 10.3969/j.issn.1003-0034.2017.04.008. Zhongguo Gu Shang. 2017. PMID: 29349981 Chinese.
-
Fracture of cementless femoral stems at the mid-stem junction in modular revision hip arthroplasty systems.J Bone Joint Surg Am. 2011 Jan 5;93(1):57-65. doi: 10.2106/JBJS.I.01589. J Bone Joint Surg Am. 2011. PMID: 21209269
-
Modification of the Unified Classification System for periprosthetic femoral fractures after hip arthroplasty.J Orthop Sci. 2018 Nov;23(6):982-986. doi: 10.1016/j.jos.2018.07.014. Epub 2018 Aug 10. J Orthop Sci. 2018. PMID: 30104103
-
Modular femoral component for conversion of previous hip surgery in total hip arthroplasty.Orthopedics. 2005 Sep;28(9 Suppl):s1079-84. doi: 10.3928/0147-7447-20050902-16. Orthopedics. 2005. PMID: 16190041 Review.
Cited by
-
Fracture of a Sivash-Range of Motion Femoral Stem at the Diaphyseal Spline.Cureus. 2021 Dec 24;13(12):e20660. doi: 10.7759/cureus.20660. eCollection 2021 Dec. Cureus. 2021. PMID: 35103211 Free PMC article.
-
Spontaneous Fractures of a Modern Modular Uncemented Femoral Stem.HSS J. 2016 Oct;12(3):250-254. doi: 10.1007/s11420-016-9510-z. Epub 2016 Jun 21. HSS J. 2016. PMID: 27703419 Free PMC article.
-
Modular prosthesis fracture in a patient with developmental dysplasia of the hip: a case report and literature review.BMC Musculoskelet Disord. 2021 May 14;22(1):442. doi: 10.1186/s12891-021-04325-2. BMC Musculoskelet Disord. 2021. PMID: 33990209 Free PMC article. Review.
-
Femoral Stem Fracture in Hip Revision Arthroplasty: A Systematic Literature Review of the Real-World Evidence.Z Orthop Unfall. 2022 Apr;160(2):160-171. doi: 10.1055/a-1348-2873. Epub 2021 Apr 13. Z Orthop Unfall. 2022. PMID: 33851402 Free PMC article.
References
-
- Adamany DC, Politi JR, Hauser WH. S-ROM hip prosthesis: 10- to 14-year results. Orthopedics 2008;31:220. - PubMed
-
- Christie MJ, DeBoer DK, Trick LW, et al. Primary total hip arthroplasty with use of the modular S-ROM prosthesis. Four to seven-year clinical and radiographic results. J Bone Joint Surg Am 1999;81:1707–16 - PubMed
-
- Skinner HB, Curlin FJ. Decreased pain with lower flexural rigidity of uncemented femoral prostheses. Orthopedics 1990;13:1222 - PubMed
-
- Cameron HU. The 3- to 6-year results of a modular noncemented low-bending stiffness hip implant: a preliminary study. J Arthroplasty 1993;8:239. - PubMed