Dose and image quality of cone-beam computed tomography as compared with conventional multislice computed tomography in abdominal imaging
- PMID: 24853071
- DOI: 10.1097/RLI.0000000000000069
Dose and image quality of cone-beam computed tomography as compared with conventional multislice computed tomography in abdominal imaging
Abstract
Objectives: Recent technical developments have facilitated the application of cone-beam computed tomography (CBCT) for interventional and intraoperative imaging. The aim of this study was to compare the radiation doses and image quality in CBCT with those of conventional multislice spiral computed tomography (MSCT) for abdominal and genitourinary imaging.
Methods: Different CBCT and MSCT protocols for imaging soft tissues and hard-contrast objects at different dose levels were investigated in this study. Local skin and organ doses were measured with thermoluminescent dosimeters placed in an anthropomorphic phantom. Moreover, the contrast-to-noise ratio, the noise-power spectrum, and the high-contrast resolution derived from the modulation transfer function were determined in a phantom with the same absorption properties as those of anthropomorphic phantom.
Results: The effective dose of the examined abdominal/genitourinary CBCT protocols ranged between 0.35 mSv and 18.1 mSv. As compared with MSCT, the local skin dose of CBCT examinations could locally reach much higher doses up to 190 mGy. The effective dose necessary to realize the same contrast-to-noise ratio with CBCT and MSCT depended on the MSCT convolution kernel: the MSCT dose was smaller than the corresponding CBCT dose for a soft kernel but higher than that for a hard kernel. The noise-power spectrum of the CBCT images at tube voltages of 85/90 kV(p) is at least half of that of images measured at 103/115 kV(p) at any arbitrarily chosen spatial frequency. Although the pixel size and slice thickness of CBCT were half of those of the MSCT images, high-contrast resolution was inferior to the MSCT images reconstructed with a hard convolution kernel.
Conclusions: As compared with MSCT using a medium-hard convolution kernel, CBCT produces images at medium noise levels and, simultaneously, medium spatial resolution at approximately the same dose. It is well suited for visualizing hard-contrast objects in the abdomen with relatively low image noise and patient dose. For the detection of low-contrast objects at standard tube voltages of approximately 120 kV(p), however, MSCT should be preferred.
Similar articles
-
Performance of cone-beam computed tomography and multidetector computed tomography in diagnostic imaging of the midface: A comparative study on Phantom and cadaver head scans.Eur Radiol. 2017 Feb;27(2):790-800. doi: 10.1007/s00330-016-4387-2. Epub 2016 May 11. Eur Radiol. 2017. PMID: 27169574
-
Dosimetry and image quality of four dental cone beam computed tomography scanners compared with multislice computed tomography scanners.Dentomaxillofac Radiol. 2009 Sep;38(6):367-78. doi: 10.1259/dmfr/15779208. Dentomaxillofac Radiol. 2009. PMID: 19700530
-
Cone beam computed tomography and low-dose multislice computed tomography in orthodontics and dentistry: a comparative evaluation on image quality and radiation exposure.J Orofac Orthop. 2014 Sep;75(5):384-98. doi: 10.1007/s00056-014-0232-x. Epub 2014 Aug 28. J Orofac Orthop. 2014. PMID: 25158951
-
Application of cone beam computed tomography for assessment of the temporomandibular joints.Aust Dent J. 2012 Mar;57 Suppl 1:109-18. doi: 10.1111/j.1834-7819.2011.01663.x. Aust Dent J. 2012. PMID: 22376103 Review.
-
[Cone Beam computerized tomography and implants].Rev Stomatol Chir Maxillofac. 2012 Sep;113(4):245-58. doi: 10.1016/j.stomax.2012.05.009. Epub 2012 Aug 11. Rev Stomatol Chir Maxillofac. 2012. PMID: 22889565 Review. French.
Cited by
-
Radiation exposure of the interventional radiologist during percutaneous biopsy using a multiaxis interventional C-arm CT system with 3D laser guidance: a phantom study.Br J Radiol. 2015;88(1055):20150151. doi: 10.1259/bjr.20150151. Epub 2015 Sep 15. Br J Radiol. 2015. PMID: 26370153 Free PMC article.
-
Postoperative computed tomography imaging of pediatric patients with craniosynostosis: radiation dose and image quality comparison between multi-slice computed tomography and O-arm cone-beam computed tomography.Pediatr Radiol. 2023 Jul;53(8):1704-1712. doi: 10.1007/s00247-023-05644-3. Epub 2023 Mar 27. Pediatr Radiol. 2023. PMID: 36967418 Free PMC article.
-
1.5 T MR-Guided Daily Adapted SBRT on Lymph Node Oligometastases from Prostate Cancer.J Clin Med. 2022 Nov 10;11(22):6658. doi: 10.3390/jcm11226658. J Clin Med. 2022. PMID: 36431135 Free PMC article.
-
Three-dimensional assessment of the favorability of maxillary posterior teeth intrusion in different facial patterns limited by the vertical relationship with the maxillary sinus floor.Clin Oral Investig. 2022 Jul;26(7):4905-4915. doi: 10.1007/s00784-022-04458-1. Epub 2022 Mar 20. Clin Oral Investig. 2022. PMID: 35306610
-
Prototype metal artefact reduction algorithm in flat panel computed tomography - evaluation in patients undergoing transarterial hepatic radioembolisation.Eur Radiol. 2018 Jan;28(1):265-273. doi: 10.1007/s00330-017-4946-1. Epub 2017 Jul 14. Eur Radiol. 2018. PMID: 28710583
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials