Surface extraction can provide a reference for micro-CT analysis of retrieved total knee implants
- PMID: 24853191
- DOI: 10.1016/j.knee.2014.04.009
Surface extraction can provide a reference for micro-CT analysis of retrieved total knee implants
Abstract
Background: Quantitative measurements of damage and wear in orthopaedic components retrieved from patients during revision surgery can provide valuable information. However, to perform these measurements there needs to be an estimate of the original, unworn geometry of the component, often requiring multiple scans of the various sizes of components that have been retrieved. The objective of this study was to determine whether the articular and backside surfaces could be independently segmented from a micro-CT reconstruction of a tibial insert, such that a tibial insert of one thickness could be used as a reference for a tibial insert of a different thickness.
Methods: New tibial inserts of a single width but with six different thicknesses were obtained and scanned with micro-CT. An automated method was developed to computationally segment the articular and backside surfaces of the components. Variability between intact and extracted components was determined.
Results: The deviations between the comparisons of the extracted surfaces (range, 0.0004 to 0.010 mm) were less (p<0.001) than the baseline deviation between the intact surfaces (range, 0.0002 to 0.053 mm).
Conclusions: An extracted surface from one insert thickness could be used to accurately represent the surface of an insert of a different thickness. This greatly enhances the feasibility of performing retrieval studies using micro-CT as a quantitative tool, by reducing the costs and time associated with acquiring, scanning, and reconstructing multiple reference tibial insert geometries.
Clinical relevance: This will add greater detail to studies of retrieved implants, to better establish how implants are functioning in vivo.
Keywords: Micro-computed tomography; Polyethylene wear; Retrieval analysis; Total knee arthroplasty.
Copyright © 2014 Elsevier B.V. All rights reserved.
Similar articles
-
Manufacturing lot affects polyethylene tibial insert volume, thickness, and surface geometry.Proc Inst Mech Eng H. 2013 Aug;227(8):884-9. doi: 10.1177/0954411913486755. Epub 2013 Apr 26. Proc Inst Mech Eng H. 2013. PMID: 23636769
-
In vitro quantification of wear in tibial inserts using microcomputed tomography.Clin Orthop Relat Res. 2011 Jan;469(1):107-12. doi: 10.1007/s11999-010-1490-6. Clin Orthop Relat Res. 2011. PMID: 20676810 Free PMC article.
-
Determination of reference geometry for polyethylene tibial insert wear analysis.J Arthroplasty. 2011 Apr;26(3):497-503. doi: 10.1016/j.arth.2010.01.096. Epub 2010 Mar 26. J Arthroplasty. 2011. PMID: 20347254
-
Tibial interface wear in retrieved total knee components and correlations with modular insert motion.J Bone Joint Surg Am. 2002 Oct;84(10):1849-55. doi: 10.2106/00004623-200210000-00017. J Bone Joint Surg Am. 2002. PMID: 12377918
-
Measurements of surface and subsurface damage in retrieved polyethylene tibial inserts of a contemporary design.J Long Term Eff Med Implants. 2012;22(1):21-31. doi: 10.1615/jlongtermeffmedimplants.v22.i1.40. J Long Term Eff Med Implants. 2012. PMID: 23016787
Cited by
-
Peripheral snap-fit locking mechanisms and smooth surface finish of tibial trays reduce backside wear in fixed-bearing total knee arthroplasty.Acta Orthop. 2017 Feb;88(1):62-69. doi: 10.1080/17453674.2016.1248202. Epub 2016 Oct 26. Acta Orthop. 2017. PMID: 27781667 Free PMC article.
-
Current Total Knee Designs: Does Baseplate Roughness or Locking Mechanism Design Affect Polyethylene Backside Wear?Clin Orthop Relat Res. 2017 Dec;475(12):2970-2980. doi: 10.1007/s11999-017-5494-3. Epub 2017 Sep 13. Clin Orthop Relat Res. 2017. PMID: 28905208 Free PMC article.
-
Relationship of surface damage appearance and volumetric wear in retrieved TKR polyethylene liners.J Biomed Mater Res B Appl Biomater. 2017 Oct;105(7):2053-2059. doi: 10.1002/jbm.b.33684. Epub 2016 Jul 12. J Biomed Mater Res B Appl Biomater. 2017. PMID: 27401236 Free PMC article.
-
Wear Distribution Detection of Knee Joint Prostheses by Means of 3D Optical Scanners.Materials (Basel). 2017 Mar 30;10(4):364. doi: 10.3390/ma10040364. Materials (Basel). 2017. PMID: 28772725 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources