A new method of sweat testing: the CF Quantum®sweat test
- PMID: 24862724
- PMCID: PMC4510978
- DOI: 10.1016/j.jcf.2014.05.001
A new method of sweat testing: the CF Quantum®sweat test
Abstract
Background: Conventional methods of sweat testing are time consuming and have many steps that can and do lead to errors. This study compares conventional sweat testing to a new quantitative method, the CF Quantum® (CFQT) sweat test. This study tests the diagnostic accuracy and analytic validity of the CFQT.
Methods: Previously diagnosed CF patients and patients who required a sweat test for clinical indications were invited to have the CFQT test performed. Both conventional sweat testing and the CFQT were performed bilaterally on the same day. Pairs of data from each test are plotted as a correlation graph and Bland-Altman plot. Sensitivity and specificity were calculated as well as the means and coefficient of variation by test and by extremity. After completing the study, subjects or their parents were asked for their preference of the CFQT and conventional sweat testing.
Results: The correlation coefficient between the CFQT and conventional sweat testing was 0.98 (95% confidence interval: 0.97-0.99). The sensitivity and specificity of the CFQT in diagnosing CF was 100% (95% confidence interval: 94-100%) and 96% (95% confidence interval: 89-99%), respectively. In one center in this three center multicenter study, there were higher sweat chloride values in patients with CF and also more tests that were invalid due to discrepant values between the two extremities. The percentage of invalid tests was higher in the CFQT method (16.5%) compared to conventional sweat testing (3.8%) (p < 0.001). In the post-test questionnaire, 88% of subjects/parents preferred the CFQT test.
Conclusions: The CFQT is a fast and simple method of quantitative sweat chloride determination. This technology requires further refinement to improve the analytic accuracy at higher sweat chloride values and to decrease the number of invalid tests.
Keywords: Cystic fibrosis; Newborn screening; Sensitivity; Specificity; Sweat testing.
Copyright © 2014 European Cystic Fibrosis Society. Published by Elsevier B.V. All rights reserved.
Conflict of interest statement
Conflict of Interest Statement
None of the authors have any personal or any financial relationships that would constitute a conflict of interest.
Figures




References
-
- LeGrys VA, Wood RE. Incidence and implications of false-negative sweat test reports in patients with cystic fibrosis. Pediatr Pulmonol. 1988;4:169–172. - PubMed
-
- Yeung WH, Palmer J, Schidlow D, et al. Evaluation of a paper-patch test for sweat chloride determination. Clin Pediatr. 1984;23:603–607. - PubMed
-
- Bossuyt PM, Reitsma JB, Bruns DE, et al. The STARD statement for reporting studies of diagnostic accuracy: explanation and elaboration. Ann Intern Med. 2003;138:W1–W12. - PubMed
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical