Radiation exposure of contrast-enhanced spectral mammography compared with full-field digital mammography
- PMID: 24872005
- DOI: 10.1097/RLI.0000000000000068
Radiation exposure of contrast-enhanced spectral mammography compared with full-field digital mammography
Abstract
Objectives: Contrast-enhanced spectral mammography (CESM) shows promising initial results but comes at the cost of increased dose as compared with full-field digital mammography (FFDM). We aimed to quantitatively assess the dose increase of CESM in comparison with FFDM.
Materials and methods: Radiation exposure-related data (such as kilovoltage, compressed breast thickness, glandularity, entrance skin air kerma (ESAK), and average glandular dose (AGD) were retrieved for 47 CESM and 715 FFDM patients. All examinations were performed on 1 mammography unit. Radiation dose values reported by the unit were validated by phantom measurements. Descriptive statistics of the patient data were generated using a statistical software package.
Results: Dose values reported by the mammography unit were in good qualitative agreement with those of phantom measurements. Mean ESAK was 10.5 mGy for a CESM exposure and 7.46 mGy for an FFDM exposure. Mean AGD for a CESM exposure was 2.80 mGy and 1.55 mGy for an FFDM exposure.
Conclusions: Compared with our institutional FFDM, the AGD of a single CESM exposure is increased by 1.25 mGy (+81%), whereas ESAK is increased by 3.07 mGy (+41%). Dose values of both techniques meet the recommendations for maximum dose in mammography.
Similar articles
-
Breast Radiation Dose With CESM Compared With 2D FFDM and 3D Tomosynthesis Mammography.AJR Am J Roentgenol. 2017 Feb;208(2):362-372. doi: 10.2214/AJR.16.16743. AJR Am J Roentgenol. 2017. PMID: 28112559
-
Comparison of radiation dose in contrast-enhanced spectral mammography (CESM), full-field digital mammography (FFDM) and digital breast tomosynthesis (DBT).Radiologia (Engl Ed). 2025 Jul-Aug;67(4):101546. doi: 10.1016/j.rxeng.2025.101546. Epub 2025 Jun 3. Radiologia (Engl Ed). 2025. PMID: 40675727
-
A survey of patient dose and clinical factors in a full-field digital mammography system.Radiat Prot Dosimetry. 2005;114(1-3):375-9. doi: 10.1093/rpd/nch514. Radiat Prot Dosimetry. 2005. PMID: 15933140
-
Current Status of Contrast Enhanced Mammography: A Comprehensive Review.Can Assoc Radiol J. 2022 Feb;73(1):141-156. doi: 10.1177/08465371211029047. Epub 2021 Sep 7. Can Assoc Radiol J. 2022. PMID: 34492211 Review.
-
Angiomammography: A review of current evidences.Diagn Interv Imaging. 2019 Oct;100(10):593-605. doi: 10.1016/j.diii.2019.01.011. Epub 2019 Apr 5. Diagn Interv Imaging. 2019. PMID: 30962168 Review.
Cited by
-
Low-Dose, Contrast-Enhanced Mammography Compared to Contrast-Enhanced Breast MRI: A Feasibility Study.J Magn Reson Imaging. 2020 Aug;52(2):589-595. doi: 10.1002/jmri.27079. Epub 2020 Feb 14. J Magn Reson Imaging. 2020. PMID: 32061002 Free PMC article.
-
Understanding the risk of ionizing radiation in breast imaging: Concepts and quantities, clinical importance, and future directions.Eur J Radiol. 2024 Dec;181:111784. doi: 10.1016/j.ejrad.2024.111784. Epub 2024 Oct 12. Eur J Radiol. 2024. PMID: 39423780 Review.
-
Empowering breast cancer diagnosis and radiology practice: advances in artificial intelligence for contrast-enhanced mammography.Front Radiol. 2024 Jan 5;3:1326831. doi: 10.3389/fradi.2023.1326831. eCollection 2023. Front Radiol. 2024. PMID: 38249158 Free PMC article. Review.
-
Contrast-enhanced spectral mammography.Singapore Med J. 2024 Mar 1;65(3):195-201. doi: 10.4103/singaporemedj.SMJ-2021-268. Epub 2024 Mar 26. Singapore Med J. 2024. PMID: 38527307 Free PMC article. No abstract available.
-
Scanning K-edge subtraction (SKES) imaging with laser-compton x-ray sources.Med Phys. 2025 Apr;52(4):2475-2492. doi: 10.1002/mp.17638. Epub 2025 Jan 28. Med Phys. 2025. PMID: 39876071 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical