Improved image quality of cone beam CT scans for radiotherapy image guidance using fiber-interspaced antiscatter grid
- PMID: 24877821
- DOI: 10.1118/1.4875978
Improved image quality of cone beam CT scans for radiotherapy image guidance using fiber-interspaced antiscatter grid
Abstract
Purpose: Medical linear accelerator mounted cone beam CT (CBCT) scanner provides useful soft tissue contrast for purposes of image guidance in radiotherapy. The presence of extensive scattered radiation has a negative effect on soft tissue visibility and uniformity of CBCT scans. Antiscatter grids (ASG) are used in the field of diagnostic radiography to mitigate the scatter. They usually do increase the contrast of the scan, but simultaneously increase the noise. Therefore, and considering other scatter mitigation mechanisms present in a CBCT scanner, the applicability of ASGs with aluminum interspacing for a wide range of imaging conditions has been inconclusive in previous studies. In recent years, grids using fiber interspacers have appeared, providing grids with higher scatter rejection while maintaining reasonable transmission of primary radiation. The purpose of this study was to evaluate the impact of one such grid on CBCT image quality.
Methods: The grid used (Philips Medical Systems) had ratio of 21:1, frequency 36 lp/cm, and nominal selectivity of 11.9. It was mounted on the kV flat panel detector of an Elekta Synergy linear accelerator and tested in a phantom and a clinical study. Due to the flex of the linac and presence of gridline artifacts an angle dependent gain correction algorithm was devised to mitigate resulting artifacts. Scan reconstruction was performed using XVI4.5 augmented with inhouse developed image lag correction and Hounsfield unit calibration. To determine the necessary parameters for Hounsfield unit calibration and software scatter correction parameters, the Catphan 600 (The Phantom Laboratory) phantom was used. Image quality parameters were evaluated using CIRS CBCT Image Quality and Electron Density Phantom (CIRS) in two different geometries: one modeling head and neck and other pelvic region. Phantoms were acquired with and without the grid and reconstructed with and without software correction which was adapted for the different acquisition scenarios. Parameters used in the phantom study were t(cup) for nonuniformity and contrast-to-noise ratio (CNR) for soft tissue visibility. Clinical scans were evaluated in an observer study in which four experienced radiotherapy technologists rated soft tissue visibility and uniformity of scans with and without the grid.
Results: The proposed angle dependent gain correction algorithm suppressed the visible ring artifacts. Grid had a beneficial impact on nonuniformity, contrast to noise ratio, and Hounsfield unit accuracy for both scanning geometries. The nonuniformity reduced by 90% for head sized object and 91% for pelvic-sized object. CNR improved compared to no corrections on average by a factor 2.8 for the head sized object, and 2.2 for the pelvic sized phantom. Grid outperformed software correction alone, but adding additional software correction to the grid was overall the best strategy. In the observer study, a significant improvement was found in both soft tissue visibility and nonuniformity of scans when grid is used.
Conclusions: The evaluated fiber-interspaced grid improved the image quality of the CBCT system for broad range of imaging conditions. Clinical scans show significant improvement in soft tissue visibility and uniformity without the need to increase the imaging dose.
Similar articles
-
Optimal combination of anti-scatter grids and software correction for CBCT imaging.Med Phys. 2017 Sep;44(9):4437-4451. doi: 10.1002/mp.12385. Epub 2017 Aug 2. Med Phys. 2017. PMID: 28556204
-
3D-printed large-area focused grid for scatter reduction in cone-beam CT.Med Phys. 2023 Jan;50(1):240-258. doi: 10.1002/mp.16005. Epub 2022 Oct 26. Med Phys. 2023. PMID: 36215176
-
Evaluation of scatter rejection and correction performance of 2D antiscatter grids in cone beam computed tomography.Med Phys. 2021 Apr;48(4):1846-1858. doi: 10.1002/mp.14756. Epub 2021 Mar 4. Med Phys. 2021. PMID: 33554377 Free PMC article.
-
A general framework and review of scatter correction methods in x-ray cone-beam computerized tomography. Part 1: Scatter compensation approaches.Med Phys. 2011 Jul;38(7):4296-311. doi: 10.1118/1.3599033. Med Phys. 2011. PMID: 21859031 Review.
-
Kilovoltage cone-beam computed tomography imaging dose estimation and optimization: Need of daily cone-beam computed tomography.J Cancer Res Ther. 2019 Jul-Sep;15(3):470-474. doi: 10.4103/jcrt.JCRT_949_17. J Cancer Res Ther. 2019. PMID: 31169206 Review.
Cited by
-
The Mechanism of Computed Tomography-Guided 125I Particle in Treating Lung Cancer.Med Sci Monit. 2017 Jan 17;23:292-299. doi: 10.12659/msm.898526. Med Sci Monit. 2017. PMID: 28095393 Free PMC article.
-
A quantitative CBCT pipeline based on 2D antiscatter grid and grid-based scatter sampling for image-guided radiation therapy.Med Phys. 2023 Dec;50(12):7980-7995. doi: 10.1002/mp.16681. Epub 2023 Sep 4. Med Phys. 2023. PMID: 37665760 Free PMC article.
-
Super-resolution CBCT on a new generation flat panel imager of a C-arm gantry linear accelerator.Med Phys. 2025 Jul;52(7):e18000. doi: 10.1002/mp.18000. Med Phys. 2025. PMID: 40665524
-
Learning-based CBCT correction using alternating random forest based on auto-context model.Med Phys. 2019 Feb;46(2):601-618. doi: 10.1002/mp.13295. Epub 2018 Dec 11. Med Phys. 2019. PMID: 30471129 Free PMC article.
-
Transmission characteristics of a two dimensional antiscatter grid prototype for CBCT.Med Phys. 2017 Aug;44(8):3952-3964. doi: 10.1002/mp.12346. Epub 2017 Jun 16. Med Phys. 2017. PMID: 28513847 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources