Closing in on the K edge: coronary CT angiography at 100, 80, and 70 kV-initial comparison of a second- versus a third-generation dual-source CT system
- PMID: 24877984
- DOI: 10.1148/radiol.14140244
Closing in on the K edge: coronary CT angiography at 100, 80, and 70 kV-initial comparison of a second- versus a third-generation dual-source CT system
Abstract
Purpose: To prospectively evaluate radiation and contrast medium requirements for performing high-pitch coronary computed tomographic (CT) angiography at 70 kV using a third-generation dual-source CT system in comparison to a second-generation dual-source CT system.
Materials and methods: All patients gave informed consent for this institutional review board-approved study. Forty-five patients (median age, 52 years; 27 men) were imaged in high-pitch mode with a third-generation dual-source CT system at 70 kV (n = 15) or with a second-generation dual-source CT system at 80 or 100 kV (n = 15 for each). Tube voltage was based on body mass index: 80 or 70 kV for less than 26 kg/m(2) versus 100 kV for 26-30 kg/m(2). For the 80- and 100-kV protocols, 80 mL of contrast material was injected, versus 45 mL for the 70-kV protocol. Data were reconstructed by using a second-generation iterative reconstruction algorithm for second-generation dual-source CT and a recently introduced third-generation iterative reconstruction algorithm for third-generation dual-source CT. Objective image quality was measured for various regions of interest, and subjective image quality was evaluated with a five-point Likert scale.
Results: The signal-to-noise ratio of the coronary CT angiography studies acquired with 70 kV was significantly higher (70 kV: 14.3-17.6 vs 80 kV: 7.1-12.9 vs 100 kV: 9.8-12.9; P < .0497) than those acquired with the other two protocols for all coronary arteries. Qualitative image quality analyses revealed no significant differences between the three CT angiography protocols (median score, 5; P > .05). The mean effective dose was 75% and 108% higher (0.92 mSv ± 0.3 [standard deviation] and 0.78 mSv ± 0.2 vs 0.44 mSv ± 0.1; P < .0001), respectively, for the 80- and 100-kV CT angiography protocols than for the 70-kV CT angiography protocol.
Conclusion: In nonobese patients, third-generation high-pitch coronary dual-source CT angiography at 70 kV results in robust image quality for studying the coronary arteries, at significantly reduced radiation dose (0.44 mSv) and contrast medium volume (45 mL), thus enabling substantial radiation dose and contrast medium savings as compared with second-generation dual-source CT.
Similar articles
-
64-slice coronary computed tomography angiography using low tube voltage of 80 kV in subjects with normal body mass indices: comparative study using 120 kV.Acta Radiol. 2012 Dec 1;53(10):1099-106. doi: 10.1258/ar.2012.110573. Epub 2012 Oct 3. Acta Radiol. 2012. PMID: 23034796
-
Thoracic aorta: prospective electrocardiographically triggered CT angiography with dual-source CT--feasibility, image quality, and dose reduction.Radiology. 2010 Apr;255(1):207-17. doi: 10.1148/radiol.09090860. Epub 2010 Feb 16. Radiology. 2010. PMID: 20160003 Clinical Trial.
-
Prospectively ECG-triggered high-pitch coronary angiography with third-generation dual-source CT at 70 kVp tube voltage: feasibility, image quality, radiation dose, and effect of iterative reconstruction.J Cardiovasc Comput Tomogr. 2014 Nov-Dec;8(6):418-25. doi: 10.1016/j.jcct.2014.09.003. Epub 2014 Sep 16. J Cardiovasc Comput Tomogr. 2014. PMID: 25439789
-
Current and Novel Imaging Techniques in Coronary CT.Radiographics. 2015 Jul-Aug;35(4):991-1010. doi: 10.1148/rg.2015140181. Epub 2015 Jun 5. Radiographics. 2015. PMID: 26046942 Review.
-
Coronary CT angiography with low radiation dose.Int J Cardiovasc Imaging. 2010 Feb;26 Suppl 1:17-25. doi: 10.1007/s10554-009-9576-5. Int J Cardiovasc Imaging. 2010. PMID: 20058080 Review.
Cited by
-
Leukocyte DNA damage after reduced and conventional absorbed radiation doses using 3rd generation dual-source CT technology.Eur J Radiol Open. 2016 Jul 6;3:134-7. doi: 10.1016/j.ejro.2016.06.001. eCollection 2016. Eur J Radiol Open. 2016. PMID: 27437433 Free PMC article. Review.
-
Low contrast medium-volume third-generation dual-source computed tomography angiography for transcatheter aortic valve replacement planning.Eur Radiol. 2017 May;27(5):1944-1953. doi: 10.1007/s00330-016-4537-6. Epub 2016 Aug 23. Eur Radiol. 2017. PMID: 27553939
-
Myocardial Coverage and Radiation Dose in Dynamic Myocardial Perfusion Imaging Using Third-Generation Dual-Source CT.Korean J Radiol. 2020 Jan;21(1):58-67. doi: 10.3348/kjr.2019.0323. Korean J Radiol. 2020. PMID: 31920029 Free PMC article.
-
The influence of image quality on diagnostic performance of a machine learning-based fractional flow reserve derived from coronary CT angiography.Eur Radiol. 2020 May;30(5):2525-2534. doi: 10.1007/s00330-019-06571-4. Epub 2020 Jan 31. Eur Radiol. 2020. PMID: 32006167
-
CT pulmonary angiography in patients with acute or chronic renal insufficiency: Evaluation of a low dose contrast material protocol.Sci Rep. 2018 Jan 31;8(1):1995. doi: 10.1038/s41598-018-20254-y. Sci Rep. 2018. PMID: 29386532 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources