Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2014 May 8;15(3):4638.
doi: 10.1120/jacmp.v15i3.4638.

Interplay effect of angular dependence and calibration field size of MapCHECK 2 on RapidArc quality assurance

Affiliations

Interplay effect of angular dependence and calibration field size of MapCHECK 2 on RapidArc quality assurance

Hosang Jin et al. J Appl Clin Med Phys. .

Abstract

The purpose of this study is to investigate an effect of angular dependence and calibration field size of MapCHECK 2 on RapidArc QA for 6, 8, 10, and 15 MV. The angular dependence was investigated by comparing MapCHECK 2 measurements in MapPHAN-MC2 to the corresponding Eclipse calculations every 10° using 10× 10 cm2 and 3 × 3 cm2 fields. Fourteen patients were selected to make RapidArc plans using the four energies, and verification plans were delivered to two phantom setups: MapCHECK 2/MapPHAN phantom (MapPHAN QA) and MapCHECK 2 on an isocentric mounting fixture (IMF QA). Migration of MapCHECK 2 on IMF was simulated by splitting arcs every 10° and displacing an isocenter of each partial arc in the Eclipse system (IMFACTUAL QA). To investigate the effect of calibration field size, MapCHECK 2 was calibrated by two field sizes (10 × 10 cm2 and 3 × 3 cm2) and applied to all QA measurements. The γ test was implemented using criteria of 1%/1 mm, 2%/2 mm, and 3%/3 mm. A mean dose of all compared points for each plan was compared with respect to a mean effective field size of the RapidArc plan. The angular dependence was considerably high at gantry angles of 90° ± 10° and 270° ± 10° (for 10 × 10/3 × 3 cm2 at 90°, 30.6% ± 6.6%/33.4%± 5.8% (6 MV), 17.3% ± 5.3%/15.0% ± 6.8% (8 MV), 8.9%± 2.9%/7.8% ± 3.2% (10 MV), and 2.2% ± 2.3%/-1.3% ± 2.6% (15 MV)). For 6 MV, the angular dependence significantly deteriorated the γ passing rate for plans of large field size in MapPHAN QA (< 90% using 3%/3 mm); however, these plans passed the γ test in IMFACTUAL QA (> 95%). The different calibration field sizes did not make any significant dose difference for both MapPHAN QA and IMFACTUAL QA. For 8, 10, and 15 MV, the angular dependence does not make any clinically meaningful impact on MapPHAN QA. Both MapPHAN QA and IMFACTUAL QA presented clinically acceptable γ passing rates using 3%/3 mm. MapPHAN QA showed better passing rates than IMFACTUAL QA for the tighter criteria. The 10 × 10 cm2 calibration showed better agreement for plans of small effective field size (< 5 × 5 cm2) in MapPHAN QA. There was no statistical difference between IMF QA and IMFACTUAL QA. In conclusion, MapPHAN QA is not recommended for plans of large field size, especially for 6 MV, and MapCHECK2 should be calibrated using a field size similar to a mean effective field size of a RapidArc plan for better agreement for IMF QA.

PubMed Disclaimer

Figures

Figure 1
Figure 1
MapCHECK 2 device set up for measuring IMRT validation plans: (a) MapPHAN QA using MapPHAN‐MC2 on an extended tennis‐racket‐type grid couch top, and (b) IMF QA using the isocentric mounting fixture (IMF).
Figure 2
Figure 2
Displacement of central axis (CAX) of MapCHECK 2 on IMF: (a) X offset (MapCHECK 2 lateral) and (b) Y offset (MapCHECK 2 longitudinal) for different collimator angles.
Figure 3
Figure 3
Percent dose difference ((MeasurementPlan)/Plan×100(%)) of MapCHECK 2 measurements with MapPHAN‐MC2. The measurements were performed for every 10° with two different field sizes (3×3cm2 and 10×10cm2) and four different energies: (a) 6 MV, (b) 8 MV, (c) 10 MV, and (d) 15 MV. Error bars represent 1 SD of compared points around isocenter (5 points for 3×3cm2 and 13 points for 10×10cm2).
Figure 4
Figure 4
Mean dose error of all compared dose points with respect to mean equivalent‐square field size of RapidArc fields in MapPHAN QA for (a) 6 MV, (b) 8 MV, (c) 10 MV, and (d) 15 MV.
Figure 5
Figure 5
Mean dose error of all compared dose points with respect to mean equivalent‐square field size of RapidArc fields in IMFACTUAL QA for (a) 6 MV, (b) 8 MV, (c) 10 MV, and (d) 15 MV.
Figure 6
Figure 6
The tongue‐and‐groove effect on the IMF QA demonstrated by continuous hot or cold spots along MLC trajectory: (a) MapCHECK 2 diode points overlaid on the HD‐MLC pattern (inner leaf width =2.5mm, outer leaf width =5mm, and diode size =0.8×0.8mm2), and (b) failed points of 15 MV IMF QA using 2%/2mm criteria for Patient #3. (Red = measurement is high; blue = calculation is high.)

References

    1. Guckenberger M, Richter A, Krieger T, Wilbert J, Baier K, Flentje M. Is a single arc sufficient in volumetric‐modulated arc therapy (VMAT) for complex‐shaped target volumes? Radiother Oncol. 2009;93(2):259–65. - PubMed
    1. Clivio A, Fogliata A, Franzetti‐Pellanda A, et al. Volumetric‐modulated arc radiotherapy for carcinomas of the anal canal: a treatment planning comparison with fixed field IMRT. Radiother Oncol. 2009;92(1):118–24. - PubMed
    1. Davidson MT, Blake SJ, Batchelar DL, Cheung P, Mah K. Assessing the role of volumetric modulated arc therapy (VMAT) relative to IMRT and helical tomotherapy in the management of localized, locally advanced, and postoperative prostate cancer. Int J Radiat Oncol Biol Phys. 2011;80(5):1550–558. - PubMed
    1. Matuszak MM, Yan D, Grills I, Martinez A. Clinical applications of volumetric modulated arc therapy. Int J Radiat Oncol Biol Phys. 2010;77(2):608–16. - PubMed
    1. Lu S, Cheng JC, Kuo S, et al. Volumetric modulated arc therapy for nasopharyngeal carcinoma: a dosimetric comparison with TomoTherapy and step‐and‐shoot IMRT Radiother Oncol. 2012;104(3):324–30. - PubMed

MeSH terms

LinkOut - more resources