Ultrafast volumetric B1 (+) mapping for improved radiofrequency shimming in 3 tesla body MRI
- PMID: 24924914
- DOI: 10.1002/jmri.24438
Ultrafast volumetric B1 (+) mapping for improved radiofrequency shimming in 3 tesla body MRI
Abstract
Purpose: To evaluate the use of the recently proposed ultrafast B1 (+) mapping approach DREAM (Dual Refocusing Echo Acquisition Mode) for a refinement of patient adaptive radiofrequency (RF) shimming.
Materials and methods: Volumetric DREAM B1 (+) calibration scans centered in the upper abdomen were acquired in 20 patients and three volunteers with written informed consent at a clinical dual source 3 Tesla (T) MR system. Based on these data, RF transmit settings were optimized by central-slice based RF-shimming (CS-RF shim) and by a refined, multi-slice adaptive approach (MS-RF shim). Simulations were performed to compare flip angle accuracy and B1 (+) homogeneity (cv = stddev/mean) achieved by CS-RF shim versus MS-RF shim for transversal and coronal slices, and for volume shimming on the spine.
Results: By MS-RF shim, mean deviation from nominal flip angle was reduced to less than 11% in all slices, all targets, and all subjects. Relative improvements in B1 (+) cv (MS-RF shim versus CS-RF) were up to 14%/39%/47% in transversal slices/coronal slices/ spine area.
Conclusion: Volumetric information about B1 (+) can be used to further improve the accuracy and homogeneity of the B1 (+) field yielding higher diagnostic confidence, and will also be of value for various quantitative methods which are sensitive to flip angle imperfections.
Keywords: 3T; B1 mapping; RF shimming; body MRI; parallel transmit.
© 2013 Wiley Periodicals, Inc.
Similar articles
-
An in vivo comparison of the DREAM sequence with current RF shim technology.MAGMA. 2015 Apr;28(2):185-94. doi: 10.1007/s10334-014-0454-3. Epub 2014 Aug 10. MAGMA. 2015. PMID: 25108702
-
Fast multistation water/fat imaging at 3T using DREAM-based RF shimming.J Magn Reson Imaging. 2015 Jul;42(1):217-23. doi: 10.1002/jmri.24775. Epub 2014 Oct 21. J Magn Reson Imaging. 2015. PMID: 25331294
-
Slice-by-slice B1+ shimming at 7 T.Magn Reson Med. 2012 Oct;68(4):1109-16. doi: 10.1002/mrm.23319. Epub 2011 Dec 27. Magn Reson Med. 2012. PMID: 22213531
-
Clinical applications of dual-channel transmit MRI: A review.J Magn Reson Imaging. 2015 Oct;42(4):855-69. doi: 10.1002/jmri.24791. Epub 2015 Apr 9. J Magn Reson Imaging. 2015. PMID: 25854179 Review.
-
Parallel transmission for ultrahigh-field imaging.NMR Biomed. 2016 Sep;29(9):1145-61. doi: 10.1002/nbm.3313. Epub 2015 May 19. NMR Biomed. 2016. PMID: 25989904 Free PMC article. Review.
Cited by
-
An in vivo comparison of the DREAM sequence with current RF shim technology.MAGMA. 2015 Apr;28(2):185-94. doi: 10.1007/s10334-014-0454-3. Epub 2014 Aug 10. MAGMA. 2015. PMID: 25108702
-
Improved Fetal Magnetic Resonance Imaging Using a Flexible Metasurface.NMR Biomed. 2025 Apr;38(4):e70016. doi: 10.1002/nbm.70016. NMR Biomed. 2025. PMID: 39999971 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical