. 2014 Sep;56(9):755-61.
doi: 10.1007/s00234-014-1390-7. Epub 2014 Jun 15.

Woven Endobridge (WEB) Device for endovascular treatment of ruptured intracranial wide-neck aneurysms: a single-center experience

Affiliations

Woven Endobridge (WEB) Device for endovascular treatment of ruptured intracranial wide-neck aneurysms: a single-center experience

Jildaz Caroff et al. Neuroradiology. 2014 Sep.

Abstract

Introduction: Endovascular treatment of ruptured wide-neck bifurcation aneurysms presents a challenge. While still under evaluation, the Woven Endobridge (WEB) aneurysm embolization system has so far shown promising results in the treatment of complex bifurcation aneurysms. We aimed to evaluate the feasibility and short-term follow-up of endovascular treatment of ruptured wide-neck aneurysm with the WEB device.

Methods: Six patients referred to our institution for acute symptomatic subarachnoid hemorrhage (SAH) and treated with the WEB device were enrolled in this study. Clinical presentations, technical details, intraoperative and postoperative complications, and outcomes were recorded. Immediate and 3-month angiographic results were also evaluated.

Results: Three middle cerebral artery (MCA) and three anterior communicating artery aneurysms were treated between 1 and 14 days after rupturing. Average dome width was 5.8 mm (range 5-7), average neck size was 4.5 mm (range 4-5), and average dome-to-neck ratio was 1.3 (range 1-1.7). The WEB system was the exclusive treatment and was successfully deployed in all cases. Per procedural thromboembolic events occurred in two cases and were treated with intra-arterial administration of antiplatelet agents without any clinical consequences. The modified Rankin Scale (mRS) score at discharge was 0 for all patients. The 3-month angiographic follow-up showed adequate occlusion in four of our six patients (67 %).

Conclusion: From this preliminary study, the high feasibility rate and lack of need for systematic antiplatelet agents favor the WEB device providing a solution for endovascular treatment of ruptured wide-neck bifurcation aneurysms during the acute phase. However, further studies are needed to evaluate the complication rate and long-term efficiency.

PubMed Disclaimer

References

    1. AJNR Am J Neuroradiol. 2014 Mar;35(3):432-8 - PubMed
    1. Neurosurgery. 2011 Oct;69(4):837-41; discussion 842 - PubMed
    1. Radiology. 2010 Sep;256(3):916-23 - PubMed
    1. AJNR Am J Neuroradiol. 2011 Mar;32(3):607-11 - PubMed
    1. J Neuroradiol. 1997 Jun;24(1):30-44 - PubMed

Publication types

LinkOut - more resources