Quality of plasma cholesterol measurements in primary care
- PMID: 2493904
- PMCID: PMC1835598
- DOI: 10.1136/bmj.298.6669.297
Quality of plasma cholesterol measurements in primary care
Abstract
Three surveys were made of the quality of plasma cholesterol measurements performed with a commercial desktop analyser (BCL Reflotron) in primary care. Each survey included three specimens, and results were received from 37, 61, and 69 participants. Although many participants obtained satisfactory results, 8.6% of the results differed by 1.0 mmol/l or more from the target values, and the overall between instrument dispersion of results was 1.3 times that between hospital laboratories. It was found that common sources of error were poor technique and the use of outdated reagent strips. Users of such instruments outside the laboratory need help and advice with training, and guidelines for this are provided. The main recommendations are that users should establish contact with a local clinical chemistry laboratory for training and support and should participate in external quality assessment schemes.
Similar articles
-
Comparison of the use of a dry chemistry analyser in primary care in Norway and the United Kingdom.Br J Gen Pract. 1993 Jan;43(366):10-4. Br J Gen Pract. 1993. PMID: 8457356 Free PMC article.
-
Improving the quality of plasma cholesterol measurements in primary care.Scand J Clin Lab Invest Suppl. 1990;198:43-8. Scand J Clin Lab Invest Suppl. 1990. PMID: 2189209 Review.
-
Reliability of the Reflotron in the determination of cholesterol.Scand J Clin Lab Invest. 1990 May;50(3):261-71. doi: 10.3109/00365519009091577. Scand J Clin Lab Invest. 1990. PMID: 2353155
-
Performance of dry-chemistry instruments in primary health care.Scand J Clin Lab Invest. 1989 Sep;49(5):483-8. doi: 10.1080/00365518909089125. Scand J Clin Lab Invest. 1989. PMID: 2595243
-
The external quality assessment schemes in Thailand.Rinsho Byori. 2002 Feb;50(2):121-5. Rinsho Byori. 2002. PMID: 11925846 Review.
Cited by
-
Comparison of the use of a dry chemistry analyser in primary care in Norway and the United Kingdom.Br J Gen Pract. 1993 Jan;43(366):10-4. Br J Gen Pract. 1993. PMID: 8457356 Free PMC article.
-
Health checks for adults.BMJ. 1990 May 19;300(6735):1318-20. doi: 10.1136/bmj.300.6735.1318. BMJ. 1990. PMID: 2369667 Free PMC article. No abstract available.
-
Screening for presymptomatic disease.J Clin Pathol. 1991 Jul;44(7):529-38. doi: 10.1136/jcp.44.7.529. J Clin Pathol. 1991. PMID: 1856283 Free PMC article. Review. No abstract available.
-
Cholesterol screening and cholesterol lowering treatment.Qual Health Care. 1993 Jun;2(2):134-7. doi: 10.1136/qshc.2.2.134. Qual Health Care. 1993. PMID: 10131634 Free PMC article. No abstract available.
-
Screening for coronary heart disease risk factors in retail pharmacies in Sheffield, 1992.J Epidemiol Community Health. 1994 Apr;48(2):178-81. doi: 10.1136/jech.48.2.178. J Epidemiol Community Health. 1994. PMID: 8189175 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical