Effect of patient centering on patient dose and image noise in chest CT
- PMID: 24951205
- DOI: 10.2214/AJR.13.12028
Effect of patient centering on patient dose and image noise in chest CT
Abstract
Objective: The objective of our study was to evaluate the effect of vertical centering on dose and image noise in chest MDCT of different-sized patients using anthropomorphic phantoms and retrospectively studying examinations of clinical patients.
Materials and methods: Three different anthropomorphic phantoms were scanned using different vertical centering (offset ± 6 cm) and were assessed with radiation dose-monitoring software. The effect of vertical positioning on the radiation dose was studied using the volume CT dose index, dose-length product, and size-specific dose estimates for different-sized phantoms. Image noise was determined from CT number histograms. Vertical positioning for chest CT examinations of 112 patients ranging from neonates to adults were retrospectively assessed.
Results: Radiation doses were highest when using the posteroanterior scout image for automatic exposure control (AEC) and when phantoms were set in the lowest table position, and radiation doses were lowest when phantoms were set in the uppermost table position. For the adult phantom, relative doses increased by 38% in the lowest table position and decreased by 23% in the highest table position. Similarly, doses for pediatric 5-year-old and newborn phantoms were 21% and 12% higher in the lowest table position and 12% and 8% lower in the highest table position, respectively. The effect decreased when a lateral scout image was used for AEC. The relative noise was lowest when the phantoms were properly centered and increased with vertical offset. In clinical patients, we observed offset with a median value varying from 25 to 35 mm below the isocenter.
Conclusion: Regardless of patient size, most patients in this study were positioned too low, which negatively affected both patient dose and image noise. Miscentering was more pronounced in smaller pediatric patients.
Keywords: chest CT; image quality; off-centering; optimization; size-specific dose estimate.
Similar articles
-
Vertical off-centering affects organ dose in chest CT: Evidence from Monte Carlo simulations in anthropomorphic phantoms.Med Phys. 2017 Nov;44(11):5697-5704. doi: 10.1002/mp.12519. Epub 2017 Sep 25. Med Phys. 2017. PMID: 28817195
-
Impact of miscentering on patient dose and image noise in x-ray CT imaging: phantom and clinical studies.Phys Med. 2012 Jul;28(3):191-9. doi: 10.1016/j.ejmp.2011.06.002. Epub 2011 Jul 8. Phys Med. 2012. PMID: 21741870
-
Effect of vertical positioning on organ dose, image noise and contrast in pediatric chest CT--phantom study.Pediatr Radiol. 2013 Jun;43(6):673-84. doi: 10.1007/s00247-012-2611-z. Epub 2013 Jan 23. Pediatr Radiol. 2013. PMID: 23341090
-
Principles of CT: radiation dose and image quality.J Nucl Med Technol. 2007 Dec;35(4):213-25; quiz 226-8. doi: 10.2967/jnmt.106.037846. Epub 2007 Nov 15. J Nucl Med Technol. 2007. PMID: 18006597 Review.
-
Increased Computed Tomography Dose Due to Miscentering With Use of Automated Tube Voltage Selection: Phantom and Patient Study.Curr Probl Diagn Radiol. 2016 Jul-Aug;45(4):265-70. doi: 10.1067/j.cpradiol.2015.11.003. Epub 2015 Dec 15. Curr Probl Diagn Radiol. 2016. PMID: 26810714 Review.
Cited by
-
How Much Is the Dose Varying between Follow-Up CT-Examinations Performed on the Same Scanner with the Same Imaging Protocol?PLoS One. 2016 Apr 6;11(4):e0152961. doi: 10.1371/journal.pone.0152961. eCollection 2016. PLoS One. 2016. PMID: 27050659 Free PMC article.
-
Asian consortium on radiation dose of pediatric cardiac CT (ASCI-REDCARD).Pediatr Radiol. 2017 Jul;47(8):899-910. doi: 10.1007/s00247-017-3847-4. Epub 2017 Apr 24. Pediatr Radiol. 2017. PMID: 28435986
-
How patient off-centering impacts organ dose and image noise in pediatric head and thoracoabdominal CT.Eur Radiol. 2019 Dec;29(12):6790-6793. doi: 10.1007/s00330-019-06330-5. Epub 2019 Jul 5. Eur Radiol. 2019. PMID: 31278575
-
Determining body height and weight from thoracic and abdominal CT localizers in pediatric and young adult patients using deep learning.Sci Rep. 2023 Nov 3;13(1):19010. doi: 10.1038/s41598-023-46080-5. Sci Rep. 2023. PMID: 37923758 Free PMC article.
-
Effect of arm position, presence of medical devices, and off-centering during acquisition of scout image on automatic tube voltage selection and current modulation in pediatric chest CT.PLoS One. 2018 Apr 17;13(4):e0195807. doi: 10.1371/journal.pone.0195807. eCollection 2018. PLoS One. 2018. PMID: 29664918 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical