Evaluation of the BD Phoenix system for identification of a wide spectrum of clinically important yeast species: a comparison with Vitek 2-YST
- PMID: 24952986
- DOI: 10.1016/j.diagmicrobio.2014.05.011
Evaluation of the BD Phoenix system for identification of a wide spectrum of clinically important yeast species: a comparison with Vitek 2-YST
Abstract
The Phoenix Yeast ID and Vitek 2-YST panels were compared using 351 molecularly identified yeast isolates. The Phoenix showed a comparable rate of correct identification for 4 common (Phoenix, 98%; Vitek, 94%) and 45 uncommon species (Phoenix, 70%; Vitek, 64%) and had a shorter mean identification time (6-7 h).
Keywords: BD Phoenix Yeast ID; Candida; Identification; Vitek 2-YST; Yeasts.
Copyright © 2014 Elsevier Inc. All rights reserved.