Automated tube voltage adaptation in head and neck computed tomography between 120 and 100 kV: effects on image quality and radiation dose
- PMID: 24961883
- DOI: 10.1007/s00234-014-1393-4
Automated tube voltage adaptation in head and neck computed tomography between 120 and 100 kV: effects on image quality and radiation dose
Abstract
Introduction: Low tube voltage allows for computed tomography (CT) imaging with increased iodine contrast at reduced radiation dose. We sought to evaluate the image quality and potential dose reduction using a combination of attenuation based tube current modulation (TCM) and automated tube voltage adaptation (TVA) between 100 and 120 kV in CT of the head and neck.
Methods: One hundred thirty consecutive patients with indication for head and neck CT were examined with a 128-slice system capable of TCM and TVA. Reference protocol was set at 120 kV. Tube voltage was reduced to 100 kV whenever proposed by automated analysis of the localizer. An additional small scan aligned to the jaw was performed at a fixed 120 kV setting. Image quality was assessed by two radiologists on a standardized Likert-scale and measurements of signal- (SNR) and contrast-to-noise ratio (CNR). Radiation dose was assessed as CTDIvol.
Results: Diagnostic image quality was excellent in both groups and did not differ significantly (p = 0.34). Image noise in the 100 kV data was increased and SNR decreased (17.8/9.6) in the jugular veins and the sternocleidomastoid muscle when compared to 120 kV (SNR 24.4/10.3), but not in fatty tissue and air. However, CNR did not differ statistically significant between 100 (23.5/14.4/9.4) and 120 kV data (24.2/15.3/8.6) while radiation dose was decreased by 7-8%.
Conclusions: TVA between 100 and 120 kV in combination with TCM led to a radiation dose reduction compared to TCM alone, while keeping CNR constant though maintaining diagnostic image quality.
Similar articles
-
Third-generation dual-source CT of the neck using automated tube voltage adaptation in combination with advanced modeled iterative reconstruction: evaluation of image quality and radiation dose.Eur Radiol. 2016 Aug;26(8):2623-31. doi: 10.1007/s00330-015-4099-z. Epub 2015 Nov 11. Eur Radiol. 2016. PMID: 26560726
-
Automated tube voltage adaptation in combination with advanced modeled iterative reconstruction in thoracoabdominal third-generation 192-slice dual-source computed tomography: effects on image quality and radiation dose.Acad Radiol. 2015 Sep;22(9):1081-7. doi: 10.1016/j.acra.2015.05.010. Epub 2015 Jul 9. Acad Radiol. 2015. PMID: 26163706
-
Automated Tube Potential Selection as a Method of Dose Reduction for CT of the Neck: First Clinical Results.AJR Am J Roentgenol. 2015 May;204(5):1049-54. doi: 10.2214/AJR.14.12975. AJR Am J Roentgenol. 2015. PMID: 25905940
-
Image quality comparison of single-energy and dual-energy computed tomography for head and neck patients: a prospective randomized study.Eur Radiol. 2022 Nov;32(11):7700-7709. doi: 10.1007/s00330-022-08689-4. Epub 2022 Apr 20. Eur Radiol. 2022. PMID: 35441839 Free PMC article. Clinical Trial.
-
Benefits of Low-Dose CT Scan of Head for Patients With Intracranial Hemorrhage.Dose Response. 2020 Mar 6;19(1):1559325820909778. doi: 10.1177/1559325820909778. eCollection 2020 Jan-Mar. Dose Response. 2020. PMID: 32214915 Free PMC article. Review.
Cited by
-
[Imaging of the head and neck region].Radiologe. 2016 Feb;56(2):181-201; quiz 202. doi: 10.1007/s00117-015-0065-7. Radiologe. 2016. PMID: 26825063 Review. German.
-
Comparison of virtual monoenergetic and polyenergetic images reconstructed from dual-layer detector CT angiography of the head and neck.Eur Radiol. 2018 Mar;28(3):1102-1110. doi: 10.1007/s00330-017-5081-8. Epub 2017 Oct 10. Eur Radiol. 2018. PMID: 29018958
-
Third-generation dual-source CT of the neck using automated tube voltage adaptation in combination with advanced modeled iterative reconstruction: evaluation of image quality and radiation dose.Eur Radiol. 2016 Aug;26(8):2623-31. doi: 10.1007/s00330-015-4099-z. Epub 2015 Nov 11. Eur Radiol. 2016. PMID: 26560726
-
Objective and subjective image quality of primary and recurrent squamous cell carcinoma on head and neck low-tube-voltage 80-kVp computed tomography.Neuroradiology. 2015 Jun;57(6):645-51. doi: 10.1007/s00234-015-1512-x. Epub 2015 Mar 26. Neuroradiology. 2015. PMID: 25808122
-
[Imaging of the head and neck region].HNO. 2016 Mar;64(3):189-209; quiz 210. doi: 10.1007/s00106-016-0140-z. HNO. 2016. PMID: 26923486 Review. German.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical