Single-source dual-energy CT angiography with reduced iodine load in patients referred for aortoiliofemoral evaluation before transcatheter aortic valve implantation: impact on image quality and radiation dose
- PMID: 24962826
- DOI: 10.1007/s00330-014-3263-1
Single-source dual-energy CT angiography with reduced iodine load in patients referred for aortoiliofemoral evaluation before transcatheter aortic valve implantation: impact on image quality and radiation dose
Abstract
Objectives: To compare image quality and radiation dose of pre-transcatheter aortic valve implantation (TAVI) aortoiliofemoral CT angiography (AICTA) provided by standard vs. dual-energy mode with reduced iodine load protocols.
Methods: One hundred and sixty-one patients underwent a two-step CTA protocol before TAVI including cardiac CTA with injection of 65 mL of iodinated contrast agent (ICA), immediately followed by AICTA. From this second acquisition, the following three different patient groups were identified: Group 1: 52 patients with standard AICTA (60 mL ICA, 100 kVp, mA automodulation); Group 2: 48 patients with dual-energy AICTA with 50 % iodine load reduction (30 mL ICA, fast kVp switching, 600 mA); Group 3: 61 patients with an identical protocol to Group 2, but exposed to 375 mA. The qualitative/subjective image quality (13-point score) and quantitative/objective image quality (contrast attenuation and image noise) were evaluated. The radiation dose was recorded.
Results: There was no significant difference in non-diagnostic images between the three protocols. Contrast attenuation, signal-to-noise ratio and contrast-to-noise ratio were significantly higher, whereas noise was significantly lower in the standard protocol (all P < 0.05). The radiation dose was lower in the dual-energy protocol at 375 mA (P < 0.05).
Conclusions: Dual-energy AICTA before TAVI results in a reduction of iodine load while maintaining sufficient diagnostic information despite increased noise.
Key points: • Dual-energy AICTA before TAVI results in a 50 % reduction of iodine load. •The reduction of iodine load maintains sufficient image quality despite increased noise. • Using 375 mA in dual-energy mode results in a reduction of radiation dose. • A high tube current setting (600 mA) should be used in overweight patients.
Similar articles
-
320-row CT transcatheter aortic valve replacement planning with a single reduced contrast media bolus injection.PLoS One. 2018 Sep 13;13(9):e0204145. doi: 10.1371/journal.pone.0204145. eCollection 2018. PLoS One. 2018. PMID: 30212567 Free PMC article.
-
Dual source multidetector CT-angiography before Transcatheter Aortic Valve Implantation (TAVI) using a high-pitch spiral acquisition mode.Eur Radiol. 2012 Jan;22(1):51-8. doi: 10.1007/s00330-011-2233-0. Epub 2011 Aug 17. Eur Radiol. 2012. PMID: 21845463
-
Using the K-edge to improve contrast conspicuity and to lower radiation dose with a 16-MDCT: a phantom and human study.J Comput Assist Tomogr. 2006 May-Jun;30(3):391-7. doi: 10.1097/00004728-200605000-00008. J Comput Assist Tomogr. 2006. PMID: 16778612
-
Role of preprocedural computed tomography in transcatheter aortic valve implantation.Rofo. 2013 Oct;185(10):941-9. Rofo. 2013. PMID: 24490256 Review.
-
Task-based measures of image quality and their relation to radiation dose and patient risk.Phys Med Biol. 2015 Jan 21;60(2):R1-75. doi: 10.1088/0031-9155/60/2/R1. Epub 2015 Jan 7. Phys Med Biol. 2015. PMID: 25564960 Free PMC article. Review.
Cited by
-
An Intra-Individual Comparison of Low-keV Photon-Counting CT versus Energy-Integrating-Detector CT Angiography of the Aorta.Diagnostics (Basel). 2023 Dec 12;13(24):3645. doi: 10.3390/diagnostics13243645. Diagnostics (Basel). 2023. PMID: 38132229 Free PMC article.
-
Low contrast media volume in pre-TAVI CT examinations.Eur Radiol. 2016 Aug;26(8):2426-35. doi: 10.1007/s00330-015-4080-x. Epub 2015 Nov 11. Eur Radiol. 2016. PMID: 26560728 Free PMC article.
-
Diagnostic Efficacy and Safety of Low-Contrast-Dose Dual-Energy CT in Patients With Renal Impairment Undergoing Transcatheter Aortic Valve Replacement.Korean J Radiol. 2024 Jul;25(7):634-643. doi: 10.3348/kjr.2023.1207. Korean J Radiol. 2024. PMID: 38942457 Free PMC article.
-
Review of Clinical Applications of Dual-Energy CT in Patients after Endovascular Aortic Repair.J Clin Med. 2023 Dec 18;12(24):7766. doi: 10.3390/jcm12247766. J Clin Med. 2023. PMID: 38137834 Free PMC article. Review.
-
Pre-TAVI imaging: an Italian survey by the CT PRotocol optimization (CT-PRO) group.Int J Cardiovasc Imaging. 2024 Apr;40(4):831-839. doi: 10.1007/s10554-024-03052-8. Epub 2024 Jan 24. Int J Cardiovasc Imaging. 2024. PMID: 38263535 Free PMC article.
References
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous