Dynamic Scheimpflug-based assessment of keratoconus and the effects of corneal cross-linking
Abstract
Purpose: To evaluate the ability of the Corvis ST (Oculus Optikgeräte GmbH, Wetzlar, Germany) to distinguish between normal and keratoconic corneas and detect the possible stiffening effect of corneal cross-linking (CXL) in vivo.
Method: A retrospective study including 86 eyes of 86 participants (27 patients with untreated keratoconus, 27 patients with CXL-treated keratoconus, and 32 normal patients). The prospective study included 9 keratoconic eyes measured 24 days before and 97 days after CXL.
Results: The following parameters from the Corvis ST showed a significant difference between keratoconic and normal eyes: intraocular pressure, corneal thickness, first applanation time, highest concavity (HC) radius, HC deformation amplitude, and HC deflection amplitude (analysis of variance, P < .0001). Receiver operating characteristic curve analysis showed the highest area under the curve for HC deflection amplitude (0.942, P < .0001). Prediction intervals for individual measurements within the two groups overlapped considerably. No significant difference was found between the untreated and CXL-treated keratoconus groups (analysis of variance, P > .05). Measurements performed before and after CXL in the same patients showed significant difference in HC deformation amplitude (P = .018), HC time (P = .018), and second applanation time (P = .027).
Conclusions: Although the corneal deformation characteristics differed between groups of keratoconic and normal eyes, this study showed that the standard parameters of the Corvis ST cannot readily be used for diagnosis of keratoconus in the individual patient or to document the effect of CXL in vivo.
Copyright 2014, SLACK Incorporated.
Comment in
-
The mystery of collagen cross-linking when it comes to in vivo biomechanical measurements.J Refract Surg. 2014 Nov;30(11):727. doi: 10.3928/1081597X-20141021-02. J Refract Surg. 2014. PMID: 25375843 No abstract available.
-
Reply: To PMID 24972407.J Refract Surg. 2014 Nov;30(11):727-8. J Refract Surg. 2014. PMID: 25469392 No abstract available.
-
Reply.J Refract Surg. 2014 Nov;30(11):727-8. J Refract Surg. 2014. PMID: 25513318 No abstract available.
Similar articles
-
Long-term Evaluation of Corneal Biomechanical Properties After Corneal Cross-linking for Keratoconus: A 4-Year Longitudinal Study.J Refract Surg. 2018 Dec 1;34(12):849-856. doi: 10.3928/1081597X-20181012-02. J Refract Surg. 2018. PMID: 30540368
-
In Vivo Early Corneal Biomechanical Changes After Corneal Cross-linking in Patients With Progressive Keratoconus.J Refract Surg. 2017 Dec 1;33(12):840-846. doi: 10.3928/1081597X-20170922-02. J Refract Surg. 2017. PMID: 29227513
-
Increased corneal hysteresis after corneal collagen crosslinking: a study based on applanation resonance technology.JAMA Ophthalmol. 2014 Dec;132(12):1426-32. doi: 10.1001/jamaophthalmol.2014.3029. JAMA Ophthalmol. 2014. PMID: 25171564
-
Corneal biomechanical property changes following corneal collagen cross-linking in keratoconus: a systematic review and meta-regression analysis.Int Ophthalmol. 2025 Jun 28;45(1):270. doi: 10.1007/s10792-025-03617-z. Int Ophthalmol. 2025. PMID: 40580225
-
The theory and art of corneal cross-linking.Indian J Ophthalmol. 2013 Aug;61(8):416-9. doi: 10.4103/0301-4738.116069. Indian J Ophthalmol. 2013. PMID: 23925326 Free PMC article. Review.
Cited by
-
Keratoconus cone location influences ocular biomechanical parameters measured by the Ocular Response Analyzer.Eye Vis (Lond). 2024 Jan 3;11(1):2. doi: 10.1186/s40662-023-00371-0. Eye Vis (Lond). 2024. PMID: 38167119 Free PMC article.
-
Quantitative assessment of corneal biomechanical changes in vivo after photorefractive intrastromal corneal cross-linking using optical coherence elastography.Quant Imaging Med Surg. 2024 Oct 1;14(10):7640-7653. doi: 10.21037/qims-24-590. Epub 2024 Sep 26. Quant Imaging Med Surg. 2024. PMID: 39429567 Free PMC article.
-
Applanation optical coherence elastography: noncontact measurement of intraocular pressure, corneal biomechanical properties, and corneal geometry with a single instrument.J Biomed Opt. 2017 Feb 1;22(2):20502. doi: 10.1117/1.JBO.22.2.020502. J Biomed Opt. 2017. PMID: 28241272 Free PMC article.
-
Correlation between corneal biomechanical and tomographic parameters in cataract patients.Int J Ophthalmol. 2025 Jul 18;18(7):1282-1293. doi: 10.18240/ijo.2025.07.11. eCollection 2025. Int J Ophthalmol. 2025. PMID: 40688785 Free PMC article.
-
Comparison of corneal biomechanics in Sjögren's syndrome and non-Sjögren's syndrome dry eyes by Scheimpflug based device.Int J Ophthalmol. 2017 May 18;10(5):711-716. doi: 10.18240/ijo.2017.05.08. eCollection 2017. Int J Ophthalmol. 2017. PMID: 28546925 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources