Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2014 Jul;41(7):071706.
doi: 10.1118/1.4884018.

On the feasibility of comprehensive high-resolution 3D remote dosimetry

Affiliations

On the feasibility of comprehensive high-resolution 3D remote dosimetry

Titania Juang et al. Med Phys. 2014 Jul.

Abstract

Purpose: This study investigates the feasibility of remote high-resolution 3D dosimetry with the PRESAGE®/Optical-CT system. In remote dosimetry, dosimeters are shipped out from a central base institution to a remote institution for irradiation, then shipped back to the base institution for subsequent readout and analysis.

Methods: Two nominally identical optical-CT scanners for 3D dosimetry were constructed and placed at the base (Duke University) and remote (Radiological Physics Center) institutions. Two formulations of PRESAGE® (SS1, SS2) radiochromic dosimeters were investigated. Higher sensitivity was expected in SS1, which had higher initiator content (0.25% bromotrichloromethane), while greater temporal stability was expected in SS2. Four unirradiated PRESAGE® dosimeters (two per formulation, cylindrical dimensions 11 cm diameter, 8.5-9.5 cm length) were imaged at the base institution, then shipped to the remote institution for planning and irradiation. Each dosimeter was irradiated with the same simple treatment plan: an isocentric 3-field "cross" arrangement of 4 × 4 cm open 6 MV beams configured as parallel opposed laterals with an anterior beam. This simple plan was amenable to accurate and repeatable setup, as well as accurate dose modeling by a commissioned treatment planning system (Pinnacle). After irradiation and subsequent (within 1 h) optical-CT readout at the remote institution, the dosimeters were shipped back to the base institution for remote dosimetry readout 3 days postirradiation. Measured on-site and remote relative 3D dose distributions were registered to the Pinnacle dose calculation, which served as the reference distribution for 3D gamma calculations with passing criteria of 5%/2 mm, 3%/3 mm, and 3%/2 mm with a 10% dose threshold. Gamma passing rates, dose profiles, and color-maps were all used to assess and compare the performance of both PRESAGE® formulations for remote dosimetry.

Results: The best agreements between the Pinnacle plan and dosimeter readout were observed in PRESAGE® formulation SS2. Under 3%/3 mm 3D gamma passing criteria, passing rates were 91.5% ± 3.6% (SS1) and 97.4% ± 2.2% (SS2) for immediate on-site dosimetry, 96.7% ± 2.4% (SS1) and 97.6% ± 0.6% (SS2) for remote dosimetry. These passing rates are well within TG119 recommendations (88%-90% passing). Under the more stringent criteria of 3%/2 mm, there is a pronounced difference [8.0 percentage points (pp)] between SS1 formulation passing rates for immediate and remote dosimetry while the SS2 formulation maintains both higher passing rates and consistency between immediate and remote results (differences ≤ 1.2 pp) at all metrics. Both PRESAGE® formulations under study maintained high linearity of dose response (R(2) > 0.996) for 1-8 Gy over 14 days with response slope consistency within 4.9% (SS1) and 6.6% (SS2), and a relative dose distribution that remained stable over time was demonstrated in the SS2 dosimeters.

Conclusions: Remote 3D dosimetry was shown to be feasible with a PRESAGE® dosimeter formulation (SS2) that exhibited relative temporal stability and high accuracy when read off-site 3 days postirradiation. Characterization of the SS2 dose response demonstrated linearity (R(2) > 0.998) over 14 days and suggests accurate readout over longer periods of time would be possible. This result provides a foundation for future investigations using remote dosimetry to study the accuracy of advanced radiation treatments. Further work is planned to characterize dosimeter reproducibility and dose response over longer periods of time.

PubMed Disclaimer

Figures

FIG. 1.
FIG. 1.
Spectrophotometer ΔOD readings from small volume samples of the SS1 PRESAGE® formulation irradiated to a dose of 3 Gy and then stored at either 3 °C or −15 °C. Error bars depict standard deviation of 2 measurements per point. After 18 days, the samples stored at 3 °C exhibited a 66% optical density increase versus a 23% increase for the samples stored at −15 °C.
FIG. 2.
FIG. 2.
Flow chart and timeline for the remote dosimetry evaluation.
FIG. 3.
FIG. 3.
(a) Transverse, (b) sagittal, and (c) coronal views of the 3-field box treatment plan (1 AP beam, 2 lateral beams) delivered to all dosimeters.
FIG. 4.
FIG. 4.
Dose response for PRESAGE® formulations SS1 and SS2. (a) Immediate postirradiation dose response. (b) Dose response 14 days postirradiation. (c) Percent change in PRESAGE® formulation dose sensitivities over 14 days.
FIG. 5.
FIG. 5.
Transverse cross sections of representative dose distributions and gamma passing maps for both PRESAGE® formulations (SS1 and SS2) studied. All gamma values ≥1 are displayed as the same color to aid visualization of failing regions. Small variations in gamma passing values may arise from minor registration or setup errors.
FIG. 6.
FIG. 6.
Sagittal cross sections of representative dose distributions and gamma passing maps for both PRESAGE® formulations (SS1 and SS2) studied. All gamma values ≥1 are displayed as the same color to aid visualization of failing regions.
FIG. 7.
FIG. 7.
Coronal cross sections of representative dose distributions and gamma passing maps for both PRESAGE® formulations (SS1 and SS2) studied. All gamma values ≥1 are displayed as the same color to aid visualization of failing regions.
FIG. 8.
FIG. 8.
A comparison of (a) preirradiation and (b) postirradiation projection images taken of a single dosimeter at both local and remote locations, and (c) their resulting ΔOD. Although a number of variables can contribute to differences in projection images between individual scanners, these differences are largely corrected out when the difference between pre- and post-irradiation images is taken in determining ΔOD. This is further discussed in Sec. III B 2 b and illustrated in Fig. 9.
FIG. 9.
FIG. 9.
(a) Lateral and AP line profiles (ΔOD/px) across both immediate and remote reconstructed dose distributions of a representative formulation SS2 dosimeter. Profile locations are indicated by the white dotted lines. (b) When ΔOD/px is normalized to the central high dose region for the same line profiles, they are essentially identical, indicating that relative dose distributions remained stable between measurements.

References

    1. Molineu A., Hernandez N., Nguyen T., Ibbott G., and Followill D., “Credentialing results from IMRT irradiations of an anthropomorphic head and neck phantom,” Med. Phys. 40(2), 022101 (8pp.) (2013).10.1118/1.4773309 - DOI - PMC - PubMed
    1. Ibbott G. S., Followill D. S., Molineu H. A., Lowenstein J. R., Alvarez P. E., and Roll J. E., “Challenges in credentialing institutions and participants in advanced technology multi-institutional clinical trials,” Int. J. Radiat. Oncol., Biol., Phys. 71(1 Suppl), S71–S75 (2008).10.1016/j.ijrobp.2007.08.083 - DOI - PMC - PubMed
    1. Molineu A., Followill D. S., Balter P. A., Hanson W. F., Gillin M. T., Huq M. S., Eisbruch A., and Ibbott G. S., “Design and implementation of an anthropomorphic quality assurance phantom for intensity-modulated radiation therapy for the Radiation Therapy Oncology Group,” Int. J. Radiat. Oncol., Biol., Phys. 63(2), 577–583 (2005).10.1016/j.ijrobp.2005.05.021 - DOI - PubMed
    1. Followill D. S., Evans D. R., Cherry C., Molineu A., Fisher G., Hanson W. F., and Ibbott G. S., “Design, development, and implementation of the Radiological Physics Center's pelvis and thorax anthropomorphic quality assurance phantoms,” Med. Phys. 34(6), 2070–2076 (2007).10.1118/1.2737158 - DOI - PubMed
    1. Oldham M., Sakhalkar H., Guo P., and Adamovics J., “An investigation of the accuracy of an IMRT dose distribution using two- and three-dimensional dosimetry techniques,” Med. Phys. 35(5), 2072–2079 (2008).10.1118/1.2899995 - DOI - PMC - PubMed

Publication types