Corrosion on the acetabular liner taper from retrieved modular metal-on-metal total hip replacements
- PMID: 24997654
- DOI: 10.1016/j.arth.2014.05.027
Corrosion on the acetabular liner taper from retrieved modular metal-on-metal total hip replacements
Abstract
Eight retrieved metal-on-metal total hip replacements displayed corrosion damage along the cobalt-chromium alloy liner taper junction with the Ti alloy acetabular shell. Scanning electron microscopy indicated the primary mechanism of corrosion to be grain boundary and associated crevice corrosion, which was likely accelerated through mechanical micromotion and galvanic corrosion resulting from dissimilar alloys. Coordinate measurements revealed up to 4.3mm(3) of the cobalt-chromium alloy taper surface was removed due to corrosion, which is comparable to previous reports of corrosion damage on head-neck tapers. The acetabular liner-shell taper appears to be an additional source of metal corrosion products in modular total hip replacements. Patients with these prostheses should be closely monitored for signs of adverse reaction towards corrosion by-products.
Keywords: Total hip arthroplasty; acetabular shell; corrosion; metal-on-metal; taper.
Copyright © 2014 Elsevier Inc. All rights reserved.
Similar articles
-
Does Surface Topography Play a Role in Taper Damage in Head-neck Modular Junctions?Clin Orthop Relat Res. 2016 Oct;474(10):2232-42. doi: 10.1007/s11999-016-4933-x. Clin Orthop Relat Res. 2016. PMID: 27339123 Free PMC article.
-
Is there material loss at the backside taper in modular CoCr acetabular liners?Clin Orthop Relat Res. 2015 Jan;473(1):275-85. doi: 10.1007/s11999-014-3982-2. Epub 2014 Oct 16. Clin Orthop Relat Res. 2015. PMID: 25318923 Free PMC article.
-
Head Taper Corrosion Causing Head Bottoming Out and Consecutive Gross Stem Taper Failure in Total Hip Arthroplasty.J Arthroplasty. 2018 Nov;33(11):3581-3590. doi: 10.1016/j.arth.2018.07.017. Epub 2018 Jul 24. J Arthroplasty. 2018. PMID: 30100136
-
Corrosion at the head-neck interface of current designs of modular femoral components: essential questions and answers relating to corrosion in modular head-neck junctions.Bone Joint J. 2016 May;98-B(5):579-84. doi: 10.1302/0301-620X.98B5.35592. Bone Joint J. 2016. PMID: 27143725 Review.
-
Systemic toxicity related to metal hip prostheses.Clin Toxicol (Phila). 2014 Sep-Oct;52(8):837-47. doi: 10.3109/15563650.2014.944977. Epub 2014 Aug 16. Clin Toxicol (Phila). 2014. PMID: 25132471 Review.
Cited by
-
Comparison of ASTM F2129 and ASTM F746 for Evaluating Crevice Corrosion.J Test Eval. 2019 Jul;47(4):2497-2511. doi: 10.1520/jte20180585. Epub 2019 Feb 19. J Test Eval. 2019. PMID: 37680964 Free PMC article.
-
Revisions of Modular Metal-on-metal THA Have a High Risk of Early Complications.Clin Orthop Relat Res. 2019 Feb;477(2):344-350. doi: 10.1097/CORR.0000000000000363. Clin Orthop Relat Res. 2019. PMID: 30794222 Free PMC article.
-
Revision for taper corrosion at the neck-body junction following total hip arthroplasty: pearls and pitfalls.Curr Rev Musculoskelet Med. 2016 Mar;9(1):75-83. doi: 10.1007/s12178-016-9322-2. Curr Rev Musculoskelet Med. 2016. PMID: 26810063 Free PMC article.
-
Clinical significance of corrosion of cemented femoral stems in metal-on-metal hips: a retrieval study.Int Orthop. 2016 Nov;40(11):2247-2254. doi: 10.1007/s00264-016-3116-4. Epub 2016 Jan 25. Int Orthop. 2016. PMID: 26810903
-
Ion release in ceramic bearings for total hip replacement: Results from an in vitro and an in vivo study.Int Orthop. 2018 Jan;42(1):65-70. doi: 10.1007/s00264-017-3568-1. Epub 2017 Jul 20. Int Orthop. 2018. PMID: 28725970
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical