Randomized Controlled Trial
. 2014 Dec;21(6):1158-67.
doi: 10.1007/s12350-014-9942-0. Epub 2014 Jul 9.

Development and validation of a patient-tailored dose regime in myocardial perfusion imaging using CZT-SPECT

Affiliations
Randomized Controlled Trial

Development and validation of a patient-tailored dose regime in myocardial perfusion imaging using CZT-SPECT

J D van Dijk et al. J Nucl Cardiol. 2014 Dec.

Abstract

Background: Guidelines for SPECT myocardial perfusion imaging (MPI) traditionally recommend a fixed tracer dose. Yet, clinical practice shows degraded image quality in heavier patients. The aim was to optimize and validate the tracer dose and scan time to obtain a constant image quality less dependent on patients' physical characteristics.

Methods: 125 patients underwent Cadmium Zinc Telluride (CZT)-SPECT stress MPI using a fixed Tc-99m-tetrofosmin tracer dose. Image quality was scored by three physicians on a 4-point grading scale and related to the number of photon counts normalized to tracer dose and scan time. Counts were correlated with various patient-specific parameters dealing with patient size and weight to find the best predicting parameter. From these data, a formula to provide constant image quality was derived, and subsequently tested in 92 new patients.

Results: Degradation in image quality and photon counts was observed for heavier patients for all patients' specific parameters (P < .01). We found body weight to be the best-predicting parameter for image quality and derived a new dose formula. After applying this new body weight-depended tracer dose and scan time in a new group, image quality was found to be constant (P > .19) in all patients.

Conclusions: Also in CZT SPECT image quality decreases with weight. The use of a tracer dose and scan time that depends linearly on patient's body weight corrected for the varying image quality in CZT-SPECT MPI. This leads to better radiation exposure justification.

PubMed Disclaimer

References

    1. J Nucl Cardiol. 2009 Nov-Dec;16(6):927-34 - PubMed
    1. J Nucl Cardiol. 2010 Dec;17(6):1009-14 - PubMed
    1. J Nucl Cardiol. 2010 Aug;17(4):709-18 - PubMed
    1. CMAJ. 2011 Mar 8;183(4):430-6 - PubMed
    1. Eur J Nucl Med Mol Imaging. 2005 Jul;32(7):855-97 - PubMed

Publication types

MeSH terms

LinkOut - more resources